![Larson Davis Spartan 821 Reference Manual Download Page 32](http://html1.mh-extra.com/html/larson-davis/spartan-821/spartan-821_reference-manual_3720611032.webp)
Spartan 821/721 Manual
Exceedance Settings
27
3.4 Exceedance Settings
FIGURE 3-3
Exceedance Card
The Exceedance Triggers settings affect the data shown on the Exceedances
card in Overall data.
•
Frequency Weight
Select the Frequency Weighting (A, C, or Z) for
the SPL trigger counters.
•
Detector
Specifies the peak detector (Impulse, Fast, or
Slow) used for exceedances.
•
Peak Frequency Weighting
Select the peak frequency weighting (A, C, or Z)
for the Peak trigger counters.
•
Exceedance Triggers
The meter has five exceedance counters: two
SPL counters and three Peak counters. Enter
levels for which the meter will trigger an
exceedance for each SPL and Peak.
For each exceedance, the meter stores a
threshold level, counter and duration. The count
is the number of times each parameter has
exceeded the preset level. The duration is the
total accumulated duration of all exceedances
for a specific parameter. The file data displays
this information in the summary tab.