SoundAdvisor Model 831C
Before You Start
7-2
Step 1
Connect preamplifier and microphone to the 831C.
For complete instructions, see
3.3 Connecting the Preamplifier to
Meter
.
Step 2
Turn meter
On
0
. For complete instructions, see
3.6 Powering On or
Off
.
Step 3
Navigate to
Tools
Setup Manager
and indicate or check all values
desired for your measurement. For complete instructions, see
6.2
Creating a Custom Setup File
.
Step 4
Calibrate the meter. For complete instructions, see
5.4 Calibrating Your
SoundAdvisor 831C
.
7.1.1 Position the 831C
The best position for sound recording is to mount the 831C on a tripod.
If held in hand, position your body away from and on an angle to the
meter to minimize interference of the sound field at the microphone
resulting from body reflections. The microphone should be angled
away from the body.
If using a free-field microphone like the 377B02, the meter should be
“pointed” at the sound source being measured.
FIGURE 7-1
Optimal Hand Held Position (Free-Field Mic)
Microphone Extension Cable
TAKE NOTE
Always position the preamplifier/microphone in a way to minimize the
effect of reflections on the sound field near the microphone.
A shielded microphone extension cable may be placed between the
meter and the preamplifier/microphone if needed for the
measurement position. No correction is necessary when using Larson
Sound Source