SoundAdvisor Model 831C
Calibrating Your SoundAdvisor 831C
5-5
Step 5
Navigate to the
Calibration Certification
tab, which shows when the meter was
last certified by the factory and when it is due for recertification.
FIGURE 5-4
Calibration Certification Tab
1.
The date of the last certification and the due date for
the next certification.
2.
The certification interval can be changed to reflect
your own certification time line.
3.
A reminder is displayed as a popup X number of days
before the interval time has expired.
Step 4
Press
Close
(
/*/
) to save your settings.
5.4 Calibrating Your SoundAdvisor 831C
Step 1
On your SoundAdvisor, press
Tools
3
Calibrate
.
Step 2
Choose your calibrator from the
Select Calibrator
drop-down menu.
LEARN MORE
Refer to the calibrator’s user guide for more information.
TAKE NOTE
For more information on the
Other
option, see
Step 3
Insert the 1/2” microphone into the 1/2” microphone opening at the bottom of
the calibrator.
Step 4
Power the calibrator on. The calibrator turns off after 1 minute. Use it within that
time, or press the power button again.
Step 5
Use the navigation keys to select
Do Calibration
. The 831C calibrates and asks if
you want to save the results.
Step 6
Select
Yes
when complete.
TAKE NOTE
If you see an error message (such as Calibration Out of Range), verify that the
dB level for your calibrator and your meter are the same, then perform the
calibration again. Be sure to power the calibrator on each time.
1.
2.
3.