Temperature setting
38
■
Temperature setting
●
Calibrating temperature sensor
The temperature sensor or temperature compensation electrode in the combination
electrode has ±1°C accuracy without calibration. You can use a known temperature
solution to calibrate the temperature sensor.
1. Insert the temperature connector into the jack of the CH2 temperature electrode
(T).
2. Immerse the electrode into the solution until the temperature sensor is
immersed.
3. Press the
key to enter the setting mode.
4. Press the
keys to select "TC" (temperature calibration setting) and then
press the
key.
5. Make sure that "ATC" is displayed, and press the
key.
The temperature setting screen appears.
6. Enter the set temperature and press the
key to confirm.
The temperature sensor is calibrated.
To return to the measurement mode, press the
key.
Note
When initializing temperature calibration data, all settings need to be initialized. Perform
initialization by referring to “ Resetting to factory default settings ” (page 41). When
initialization is performed, all saved data is deleted. Copy or transfer necessary data to a
PC for storage.