Connecting an electrode
To perform measurement, it is necessary to use the appropriate electrode for
measurement parameters. Recommended electrodes for each measured sample are
listed in our product catalog and on our website. Refer to them when preparing
electrodes. Use the following procedure to correctly connect the electrode to the
instrument.
1. Insert the electrode connector by fitting
its groove with the connector socket pin
of the CH2 measurement electrode (G).
2. Turn the electrode connector clockwise
by following the groove.
3. Put the connector cover on the
connector.
4. When using a combination electrode
equipped with a temperature sensor,
insert the temperature connector into
the jack of the CH2 temperature
electrode (T).
Tip
If the temperature connector is unconnected or the connection is wrong, temperature
selected for MTC (manual temperature compensation) is displayed as the sample
temperature.
Preparation for measurement is complete.