Installation
14
Ensuring the System's Safe Operation
14.1
Operation with the SUH 106 Safety Enclosure
14.1.1
The user may be at risk when working on the scanner whilst this is in operation.
The safety enclosure protects the user and prevents any work on the scanner whilst this is in
operation.
Install the SUH 106 safety enclosure according to the corresponding user manual and connect the
SUH 106 connecting cable to the "ext. Stop" port of the FLS 106 scanner.
The EMERGENCY STOP switch on the scanner has to be released and the SUH 106 door closed
before the scanner can be started.
Operation with Spatial Separation and External NA 5 EMERGENCY STOP Switch
14.1.2
The user may be at risk when working on the scanner whilst this is in operation.
A spatial separation protects the user and can prevent any work on the scanner whilst this is in
operation.
The spatial separation must meet the following requirements:
- stable overall structure made of a durable material
- clearly recognizable as a separation (use bright colours if necessary)
- the separation must be provided around the scanner without any interruptions
- the distance to the scanner must be sufficient to safely prevent any interference with the scanner.
The spatial separation can be provided by different means such as separators, barriers with
barricade tape, protective fences, light barriers, light grids, etc.
The external NA 5 EMERGENCY STOP switch has to be installed outside the spatial separation so
that it can be seen and reached at any time.
The D-Sub connector of the external EMERGENCY STOP switch has to be installed before putting
the scanner into service as otherwise a fault message appears and the operation is prevented.