4
HANDLING & PRESERVATION
Samples may be stored and refrigerated in plastic containers.
INTERFERENCES
The following substances produce a positive interference at the concentration given:
Chloride (Cl-)
7000 mg/L
Phosphate (PO
4
-3
)
16 mg/L
Hexametaphosphate (NaPO
3
)
6
1 mg/L
The following substances produce a negative interference at the concentration given:
Alkalinity (CaCO
3
)
5000 mg/L
Aluminum (Al
+3
)
0.1 mg/L
Iron (Fe
+3
)
10 mg/L
Sulfate (SO
4
-2
)
200 mg/L
Color and turbidity must be removed or compensated for in the procedure. Temperature
should be maintained within 5 degrees Celcius of room temperature.
ANALYSIS
PROCEDURE
1. Fill the Water Sample Collecting
Bottle (0688) with sample water.
2. Press and hold
until
colorimeter turns on.
Main Menu
Testing Menu
Editing Menu
12:00:00 001/500
3. Press
ENTER
to select
Testing Menu
.
All Tests
045 Fluoride
12:00:00 001/500