battery charge more quickly than storage at room temperature. If the meter does not 

turn on, it means that the battery is at a very low charge. Charging the battery with the 

wall adapter in this state may take up to 10 hours. At low temperatures, approaching 0 

°C, the battery will charge more slowly. It will not charge at all below 0 °C.
The battery icon will show no bars and continuously flash if the battery is getting low 

but the unit will still operate normally. A “Low Battery” message on the status bar of the 

display will replace the time when the battery voltage is too low for proper operation 

and accuracy may be degraded. A “Shutdown Low Batt” message on the display will 

appear for a few seconds before the power is switched off when the battery is too low 

to operate the unit. When the battery icon simultaneously flashes bars 1 and 2 followed 

by bars 3 and 4, it is an indication that the battery is damaged and technical support 

should be contacted.
 To extend the life of the battery:

•  Shut down the unit with the power switch when not taking measurements or use 

the power save option to have the unit automatically turn off after 5 minutes.

•  Store the unit at a moderate temperature.

•  Fully charge the battery before storing the unit for extended periods of time. 

•  Fully charge the battery at least once per year. Failure to do so may result in a 

permanently drained battery.

•  Limit backlight use. The unit consumes three times the normal power when the 

backlight is on. Set the backlight time option to 10 seconds or select “Button 

Control” and keep the backlight off.

 

ƒ

BATTERY REPLACEMENT

The lithium ion battery used in this unit should last for many years with normal use. 

When it no longer powers the unit long enough to meet testing requirements it should 

be replaced. Contact LaMotte Company by phone (1-800-344-3100) or email (tech@

lamotte.com) for more information.

MAINTENANCE

 

„

CLEANING

Clean the exterior housing with a damp, lint-free cloth. Do not allow water to enter the 

light chamber or any other parts of the meter. To clean the light chamber and optics 

area, point a can of compressed air into the light chamber and blow the pressurized air 

into the light chamber. Use a cotton swab dampened with Windex

®

 window cleaner to 

gently swab the interior of the chamber. Do not use alcohol; it will leave a thin residue 

over the optics when dry.

 

„

RETURNS

Should it be necessary to return the meter, pack the meter carefully in a suitable 

container with adequate packing material. A return authorization number must be 

obtained from LaMotte Company by calling 800-344-3100 (US only) or 410-778-3100, 

faxing 410-778-6394, or emailing [email protected]. Often a problem can be resolved 

over the phone or by email. If a return of the meter is necessary, attach a letter with the 

return authorization number, meter serial number, a brief description of problem and 

contact information including phone and FAX numbers to the shipping carton.

 

„

METER DISPOSAL

Waste Electrical and Electronic Equipment (WEEE)
Natural resources were used in the production of this equipment. This equipment may 

contain materials that are hazardous to health and the environment. To avoid harm to 

Summary of Contents for DC1500 FLUORIDE

Page 1: ...DC1500 FLUORIDE Fluoride Colorimeter Kit Code 3243 ACCEPTED...

Page 2: ...r Connection Output 15 Computer Connection 15 WaterLink Connect 2 15 Battery Battery AC Operation 15 Battery Replacement 16 Maintenance Cleaning 16 Returns 16 Meter Disposal 16 General Operating Infor...

Page 3: ...1720 01 that is supplied with the kit Make no substitutions Test Tubes with Caps 0290 6 Replacement Chamber 3 0038 USB Cable 1720 01 USB Wall Adapter 1721 Car Charger 5 0132 TEST METHODS SPECIFICATIO...

Page 4: ...the concentration given Alkalinity CaCO3 5000 mg L Aluminum Al 3 0 1 mg L Iron Fe 3 10 mg L Sulfate SO4 2 200 mg L Color and turbidity must be removed or compensated for in the procedure Temperature s...

Page 5: ...ear colorless Fluoride free water 6 Use the 0 5 mL pipet 0353 to add 0 5 mL of Sodium Arsenite Solution 4128 Cap and mix 7 Use the 1 0 mL pipet 0354 to add 2 measures of Acid Zirconyl SPADNS Reagent 3...

Page 6: ...t blank NOTE The meter will remember the last scanned blank reading It is not necessary to scan a blank each time the test is performed To use the previous blank reading instead of scanning a new one...

Page 7: ...select TestingMenu All Tests 045 Fluoride 12 00 00 001 500 3 Scroll to and select 045 Fluoride from menu 045 Fluoride Scan Blank Scan Sample 12 00 00 001 500 4 This test requires a reagent blank Rins...

Page 8: ...imeter tube 0290 with sample water Fill to the 10 mL line with sample water 9 Use the 0 5 mL pipet 0353 to add 0 5 mL of Sodium Arsenite Solution 4128 Cap and mix 10 Use the 1 0 mL pipet 0354 to add 2...

Page 9: ...he value shown to the concentration of the prepared standard 1 0 in this example NOTE A maximum adjustment of 25 is possible 045 Fluoride 1 0 ppm Edit ENTER Save ENTER Default 12 00 00 001 500 15 Pres...

Page 10: ...ach Set Time Year 2000 12 00 00 001 500 3 Press ENTER to select the final character The time and date will be saved and the meter will return to the Editing Menu Editing Menu Set Clock Logging Factory...

Page 11: ...Editing Menu Set Clock Logging Factory Setup Set Power Save 12 00 00 001 500 FACTORY SETUP The Factory Setup menu is used in manufacturing of the colorimeter This menu is not for use by the operator i...

Page 12: ...iewing The default setting is 10 seconds If Button Control is chosen the backlight button on the key pad will act as an on off switch and the backlight will remain on or off when the meter is being us...

Page 13: ...d the meter will return to the Editing Menu Editing Menu Logging Factory Setup Set Power Save Set Backlight Time 12 00 00 001 500 SELECTING A LANGUAGE There are eight languages available English Spani...

Page 14: ...incorporate a looping feature which allows the user to quickly reach the last choice in the menu from the first choice In a looping menu the last choices in the menu are above the first choice and scr...

Page 15: ...B mini B connector into the meter and the larger end of the USB cable USB type A connector into the wall adapter Plug the wall adapter into an AC outlet Reinsert the rubber USB port plug after chargin...

Page 16: ...the backlight time option to 10 seconds or select Button Control and keep the backlight off BATTERY REPLACEMENT The lithium ion battery used in this unit should last for many years with normal use Whe...

Page 17: ...with an integrated interference filter The interference filter selects a narrow band of light from the corresponding LED for the colorimetric measurements A USB wall adapter USB computer connection o...

Page 18: ...t the top of the menu will lead to the bottom of the menu Header Menu Title Main Window Selection First Choice Second Choice Third Choice Another Footer 12 00 00 001 500 P And Another And So On A ligh...

Page 19: ...prevent dust from entering the tubes Dry tubes should be stored with the caps on to prevent contamination After a tube has been filled and capped it should be held by the cap and the outside surface s...

Page 20: ...mpany to receive compensation for damaged goods GENERAL PRECAUTIONS READTHEINSTRUCTIONMANUALBEFOREATTEMPTINGTOSETUPOROPERATETHE METER Failure to do so could result in personal injury or damage to the...

Page 21: ...RH non condensing USB Port Mini B Power Requirements USB wall adapter USB computer connection or lithium ion rechargeable battery Battery Charge Life Approximately 380 tests with backlight on to 1000...

Page 22: ...no other warranty whether written or oral is expressed or implied TROUBLESHOOTING ERROR MESSAGES Over Range If the message OVERRANGE is displayed when scanning a sample the sample may be over range o...

Page 23: ...B wall computer adapter Shut Down Low Batt Shut Down Battery is too low to operate the unit Charge battery or use USB wall computer adapter Overrange Sample is outside of acceptable range Dilute sampl...

Page 24: ...gton Ave Chestertown Maryland 21620 USA 1 410 778 3100 800 344 3100 www lamotte com LaMotte and WaterLink are registered trademarks of LaMotte Company 2023 LaMotte Company All Rights Reserved 3243 MN...

Reviews: