58
c
HAPTER
4:
Operation
Model 336 Temperature Controller
During normal operation, only the curves that share the input type you have selected
are displayed. If the curve you wish to select does not appear in the selection
sequence make sure the curve format matches the recommended format for the
input type selected. Refer to TABLE 4-7.
The sensor reading of the instrument can always be displayed in sensor units. If a tem-
perature response curve is selected for an input, its readings may also be displayed in
temperature.
Once the input is configured (section 4.4), you may choose a temperature curve. Any
standard or user curve that matches the format of the sensor type configured for a
given input will be available under the Curve parameter in the Input Setup menu for
that input. You are also given the choice of None. When set to None, front panel read-
ings configured for kelvin or Celsius will display the NOCURV message and the inter-
face will report 0 K and –273.15 °C for KRDG and CRDG queries, respectively. Data
points for standard curves are detailed in Appendix C.
Menu Navigation:
Input Setup
Q
Input
(A, B, C or D)
Q
Curve
Q
(Any curve of matching type)
4.4.10 Filter
The reading filter applies exponential smoothing to the sensor input readings. If the
filter is turned on for a sensor input, all reading values for that input are filtered. The
filter is a running average so it does not change the update rate of an input. Filtered
readings are not used for control functions but they are used for all input features
including Max/Min.
The number of filter points determines filter bandwidth. One filter point corresponds
to one new reading on that input. A larger number of points does more smoothing,
but also slows the instruments response to real changes in temperature. The default
number of filter points is 8, which settles to within six time constants of a step
change value in 45 readings, or 4.5 s.
Curve
Number
Curve Name
Sensor Type
Model Number
Temperature Range**
For Data Points, Refer To:
01
DT-470
Diode
DT-470
1.4 - 475 K
Table D-1
02
DT-670
Diode
DT-670
1.4 - 500 K
Table D-2
03
DT-500-D*
Diode
DT-500-D
1.4 - 365 K
Table D-3
04
DT-500-E1*
Diode
DT-500-E1
1.1 - 330 K
Table D-3
05
Reserved
—
—
—
—
06
PT-100
PTC RTD
PT-100
30 - 800 K
Table D-4
07
PT-1000*
PTC RTD
PT-1000
30 - 800 K
Table D-4
08
RX-102A-AA
NTC RTD
Rox RX-102A
0.05 - 40 K
Table D-5
09
RX-202A-AA
NTC RTD
Rox RX-202A
0.05 - 40 K
Table D-6
10
Reserved
—
—
—
—
11
Reserved
—
—
—
—
12
Type K
Thermocouple
Type K
3 - 1645 K
Table D-7
13
Type E
Thermocouple
Type E
3 - 1274 K
Table D-8
14
Type T
Thermocouple
Type T
3 - 670 K
Table D-9
15
AuFe 0.03%*
Thermocouple
AuFe 0.03%*
3.5 - 500 K
Table D-10
16
AuFe 0.07%
Thermocouple
AuFe 0.07%
3.15 - 610 K
Table D-11
17
Reserved
—
—
—
—
18
Reserved
—
—
—
—
19
Reserved
—
—
—
—
20
Reserved
—
—
—
—
21 - 59
User Curves
—
—
—
—
*No longer offered by Lake Shore
**Instrument may not support the sensor over its entire range
TABLE 4-11
Sensor curves
Summary of Contents for 336
Page 4: ...Model 336 Temperature Controller...
Page 6: ...Model 336 Temperature Controller...
Page 26: ...14 cHAPTER 1 Introduction Model 336 Temperature Controller...
Page 54: ...42 cHAPTER 3 Installation Model 336 Temperature Controller...
Page 84: ...72 cHAPTER 4 Operation Model 336 Temperature Controller...
Page 104: ...92 cHAPTER 5 Advanced Operation Model 336 Temperature Controller...
Page 164: ...152 cHAPTER 7 Options and Accessories Model 336 Temperature Controller...
Page 178: ...166 cHAPTER 8 Service Model 336 Temperature Controller...