154
c
HAPTER
6:
Computer Interface Operation
Model 350 Temperature Controller
RANGE
Heater Range Command
Input
RANGE <output>,<range>[term]
Format
n,n
<output>
Specifies which output to configure: 1–4.
<range>
For outputs 1 and 2: 0 = Off, 1 = Range 1, 2 = Range 2,
3 = Range 3, 4 = Range 4, 5 = Range 5
For outputs 3 and 4: 0 = Off, 1 = On
Remarks
The range setting has no effect if an output is in the Off mode, and does not apply to
an output in Monitor Out mode. An output in Monitor Out mode is always on.
RANGE?
Heater Range Query
Input
RANGE? <output>[term]
Format
n
<output>
Specifies which output to query: 1–4.
Returned
<range>[term]
Format
n (refer to command for description)
RDGST?
Input Reading Status Query
Input
RDGST? <input>[term]
Format
a
<input>
Specifies which input to query: A - D
(D1 - D5 for 3062 option).
Returned
<status bit weighting>[term]
Format
nnn
Remarks
The integer returned represents the sum of the bit weighting of the input status flag
bits. A “000” response indicates a valid reading is present..
Bit
Bit Weighting
Status Indicator
0
1
invalid reading
4
16
temp underrange
5
32
temp overrange
6
64
sensor units zero
7
128
sensor units overrange
RELAY
Relay Control Parameter Command
Input
RELAY <relay number>,<mode>,<input alarm>,<alarm type>[term]
Format
n,n,a,n
<relay number>
Specifies which relay to configure: 1 or 2.
<mode>
Specifies relay mode. 0 = Off, 1 = On, 2 = Alarms.
<input alarm>
Specifies which input alarm activates the relay when the
relay is in alarm mode: A - D (D1 - D5 for 3062 option).
<alarm type>
Specifies the input alarm type that activates the relay when
the relay is in alarm mode. 0 = Low alarm, 1 = High Alarm,
2 = Both Alarms.
Example
RELAY 1,2,B,0[term]
–relay 1 activates when Input B low alarm activates.
RELAY?
Relay Control Parameter Query
Input
RELAY? <relay number>[term]
Format
n
<relay number>
Specifies which relay to query: 1 or 2.
Returned
<mode>,<input alarm>,<alarm type>[term]
Format
n,a,n (refer to command for description)
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...