144
c
HAPTER
6:
Computer Interface Operation
Model 350 Temperature Controller
DFLT
Factory Defaults Command
Input
DFLT 99[term]
Remarks
Sets all configuration values to factory defaults and resets the instrument. The “99” is
included to prevent accidentally setting the unit to defaults.
DIOCUR
Diode Excitation Current Parameter Command
Input
DIOCUR <input>,<excitation>[term]
Format
a,n
<input>
Specifies which input to configure: D2–D5 (only for the 3062 card).
<excitation > Specifies the Diode excitation current: 0 = 10 µA, 1 = 1 mA.
Remarks
The 10 µA excitation current is the only calibrated excitation current, and is used in
almost all applications. Therefore the Model 350 will default the 10 µA current set-
ting any time the input sensor type is changed in order to prevent an accidental
change. If using a current that is not 10 µA, the input sensor type must first be config-
ured to Diode (INTYPE command). If the sensor type is not set to Diode when the
DIOCUR command is sent, the command will be ignored.
DISPFLD
Custom Mode Display Field Command
Input
DISPFLD <field>,<input>,<units>[term]
Format
n,n,n
<field>
Specifies field (display location) to configure: 1–8.
<input>
Specifies item to display in the field: 0 = None, 1 = Input A,
2 = Input B, 3 = Input C, 4 = Input D (5 = Input D2, 6 = Input D3,
7 = Input D4, 8 = Input D5 for 3062 option)
<units>
Valid entries: 1 = kelvin, 2 = Celsius, 3 = sensor units,
4 = minimum data, and 5 = maximum data.
Example
DISPFLD 2,1,1[term]
—displays kelvin reading for Input A in display field 2 when dis-
play mode is set to Custom.
Remarks
This command only applies to the readings displayed in the Custom display mode. All
other display modes have predefined readings in predefined locations, and will use
the Preferred Units parameter to determine which units to display for each sensor
input. Refer to section 4.3 for details on display setup
DISPFLD?
Custom Mode Display Field Query
Input
DISPFLD? <field>[term]
Format
n
<field>
Specifies field (display location) to query: 1–8.
Returned
<input>,<units>[term]
Format
n,n (refer to command for description)
Summary of Contents for 350
Page 4: ...Model 350 Temperature Controller ...
Page 6: ......
Page 14: ...Model 350 Temperature Controller ...
Page 28: ...14 cHAPTER 1 Introduction Model 350 Temperature Controller ...
Page 53: ...Examples of PID control 2 13 4 ManualOutput 39 www lakeshore com FIGURE 2 4 ...
Page 58: ...44 cHAPTER 2 Cooling System Design and Temperature Control Model 350 Temperature Controller ...
Page 118: ...104 cHAPTER 5 Advanced Operation Model 350 Temperature Controller ...
Page 178: ...164 cHAPTER 7 Options and Accessories Model 350 Temperature Controller ...
Page 192: ...178 cHAPTER 8 Service Model 350 Temperature Controller ...