48
Sleep Screen
Touch panel goes to sleep when the screen has been inactive for 3 minutes and shows coolant
supply temperature. The system doesn’t go to sleep screen when the information screen is being
displayed or when the unit is not running.
Figure 39 Sleep Screen
Information Screen
Figure 40 Information Screen
In order to see information related to Coolant Supply and Refrigerant, click on the Information Menu.
According to the units selected in the settings menu, the Temperature unit and Pressure unit will be
shown at the bottom.
The following information is displayed:
Information
Description
C
ool
ant
Sup
pl
y
Temperature
This is the Supply Temperature from the chiller to the application
Return Temperature
This is the Return Temperature from the application to the chiller
Pressure
This is the Supply Pressure to your application from the chiller
Temperature setpoint
This is the Temperature Setpoint set by User in the Main Menu
R
ef
rig
er
ant
Suction Temperature
This is the temperature of the refrigerant
Suction side pressure This is the pressure of the refrigerant before entering the
compressor
Condenser side
pressure
This is the pressure of the refrigerant after passing through the
condenser
Summary of Contents for Nextreme Series
Page 1: ...Nextreme Recirculating Chiller Specification and User Manual Part Number 387004848 ...
Page 17: ...17 Component Locations Figure 2 Isometric Views of Unit Figure 3 Top View ...
Page 18: ...18 Figure 4 Front View Figure 5 Rear View Figure 6 Side View ...
Page 39: ...39 Figure 25 Sensor Wiring Low Voltage ...
Page 40: ...40 Plumbing Refrigeration Diagram Figure 26 Plumbing and Refrigeration Diagram ...
Page 43: ...43 Figure 31 Pump Performance of NRC5000 with Cooling Fluids Water and 40 EG W ...
Page 60: ...60 PCFLT Returns FLT fault or NO FAULTS FOUND PCFLT C Clears Fault Returns FLT Faults Cleared ...
Page 69: ...69 Declaration of Conformity ...