laerdal SimJunior Directions For Use Manual Download Page 1

   SimJunior

Directions for Use

www.laerdal.com

Summary of Contents for SimJunior

Page 1: ...SimJunior Directions for Use www laerdal com...

Page 2: ......

Page 3: ...t Settings 15 Help Files 15 Manikin Setup 16 Preparing for Simulations 16 Using an External Compressor 16 Preparing the IO Leg 16 Administering IV Fluids 16 Draining Excess IV Fluid 17 Cleaning the IV...

Page 4: ...two conditions 1 This device may not cause harmful interference 2 This device must accept any interference received including interference that may cause undesired operation Caution Changes or modific...

Page 5: ...onal injury or damage to the product Introduce fluids into the manikin only as directed in this document as this may damage the manikin and its components Lubricate the oral and nasal airways with the...

Page 6: ...kage or mechanical damage There are signs of electrical malfunction such as an unresponsive manikin or unusual smell or smoke Warning Avoid pinch hazards Do not use the manikin without the external sk...

Page 7: ...ions 5 Introduction Right Arm ECG Connector Right Leg ECG Connector Left Arm ECG Connector Left Leg ECG Connector Power Cable 37 Pin Serial Cable Manikin toManikin Cable Adapter External Air Connectio...

Page 8: ...hing label Shirt Shorts Boxer Shorts General Clinical Features Airway Features The airway is anatomically correct to the trachea The airway can be manipulated by Head tilt chin lift Jaw thrust with ar...

Page 9: ...P Pulse palpation is detected and logged CPR Features Compliant with 2010 Guidelines Compressions generate palpable pulses blood pressure wave form and ECG artifacts Realistic compression depth and re...

Page 10: ...re cuff to the SimPad Blood Pressure connector on the back of the Link Box 3 Connect the Manikin to a power supply Network Cable Optional LAERDAL Setup of SimJunior Components 4 Refer to the SimPad DF...

Page 11: ...e that the SimJunior simulator icon is displayed on the Instructor PC desktop 2 Install the USB hub Connect Power to the USB hub Connect USB hub to Instructor PC 3 Connect the Patient Monitor Cables P...

Page 12: ...nikin Setup Instructor Software License Instructor PC is shipped from Laerdal Medical with a pre activated license installed Changes or updates of the computer hardware e g new hard drive or mother bo...

Page 13: ...n on off and pulse function central radial and brachial Set tongue edema and stomach distention Note View and configure additional sounds with the Edit menu located at the top of the screen 1 12 13 2...

Page 14: ...lick Play button to start Note Pause button changes to Play button when the scenario is paused The Pause Play button is located below the Start Scenario button When a simulation is running click this...

Page 15: ...have been included in the current scenario Select the Media button to open display Select the Close button to close the Media display Select the Labs button to open Select the Close button to close th...

Page 16: ...e available in the Laerdal DebriefViewer Note It is important to save the debrief files for later review 2 From Windows Start menu Click the Windows Start button Select All Programs Select the Laerdal...

Page 17: ...ult for future sims Internal Compressor You can configure the RPM setting for the internal compressor From the Instructor Application 1 Select the Edit menu in the Instructor Application 2 Select Conf...

Page 18: ...th IV tubing and one IV bag for use as a reservoir during infusion Additional equipment needed IV bag and administration set Simulated blood concentrate 35cc syringe and 16 gauge intraosseous needle T...

Page 19: ...ng live defibrillation the defibrillator and manikin may present a shock hazard All standard safety precautions must be taken when using the defibrillator on the manikin Caution The manikin must not b...

Page 20: ...l inserts constricted and dilated comes with SimJunior To change the pupils 1 Carefully open the eyelids wide taking care not to tear the manikin face skin 2 Using the suction cup tool provided in the...

Page 21: ...s or linens Remove tape residue with a moist cloth or rubbing alcohol Based on the use of the manikin replace modules that are consumed or damaged Replace cables or connectors showing visible damage S...

Page 22: ...ect the Defibrillation connector cable 4 The Defibrillation connector cable must be disconnected at the right side of the manikin below the chest form Note DO NOT g n i t c e n n o c s e l b a c d n a...

Page 23: ...g theTorso 2 Disconnect the clear tubing from the lung bladder and remove the lung bladder For details on removing the lung bladder see the section Replacing the Lung Bladder 3 Disconnect the clear tu...

Page 24: ...orso section 1 Remove old arm by unscrewing the bolt holding the arm to the manikin s shoulder 2 Ensure that the shoulder screw is loose enough to allow the arm axle to slide out easily 3 Discard the...

Page 25: ...e of the oral cavity with manikin wipes Change the lung bladder see section Maintenance Replacing Lung Bladder Chest Movement Problem No chest rise on manikin Possible Solutions Check that power to ma...

Page 26: ...ial and Brachial Problem Cannot feel pulses Possible Solutions Skin may be too tight over pulse units re adjust skin and reboot Ensure blood pressure is set to at least 60 systolic Ensure compressor i...

Page 27: ...25 Notes...

Page 28: ...0 ft max Material Chart for Manikin Clothes Cotton Nylon Skins and airways PVC External hard plastics PVC ABS Inner plastics Silicone TPU PVC Nitrile ABS POM Nylon GF Epoxy Polyurethane Metal componen...

Page 29: ...27 Notes...

Page 30: ...Chest Rise Bladder 375 70150 Pediatric Skin Vein Set Light 200 03050 Pupil Inserts Kit Blue 200 03050B Pupil Inserts Kit Brown 375 71001 Right IV Arm 232 01150 Left BP Arm 232 00450 BP Arm Skin 231 0...

Page 31: ...29 Notes...

Page 32: ......

Page 33: ......

Page 34: ...2012 Laerdal Medical All rights reserved 20 05754 Rev A Printed in USA...