www.laerdal.com
EN
50
80
O
2
%
O
2
flow
54
40
5
10
15
N
O
SE
N
O
SE
1
2
1
2
N
O
SE
Mask Filter Valve
Oxygen
inlet
Hard
case
Soft
Pouch
Poly
Bag
PVC PP
PC and
sillicone
rubber
PC and
sillicone
rubber
PP
PA
LDPE
Cleaning and Disinfection
Warning
The One-way Valve and Filter are designed for
single patient use only. Do not re-use. Re-use
will lead to increased risk of cross contamination,
degradation of performance, and/or device
malfunction. Laerdal is not responsible for any
consequences of re-use.
Mask can be cleaned and disinfected, and
reused after manikin training and patient use.
1. Wash and scrub in warm soapy water.
2. Rinse in clean water for min 1 min.
3. Submerge for 10 minutes in a 0.5%
Sodium Hypochlorite solution.
4. Rinse under running tap water for 3 min.
5. Allow to dry completely.
Patient Use
One-way Valve and filter must be discarded
after single patient use.
After training on a manikin
The One-way Valve should be cleaned using
the above procedure after each user session.
Cautions
• Filter must be discarded after use.
• Do not pasteurize, boil or steam autoclave.
Not made with natural rubber latex
Use of Mask with Oxygen
Remove Used Filter
Hold Mask and pull Filter outwards and
sideways, so it separates from Mask.
Place the mask on a hard, flat surface and
insert a new filter. Firm pressure is required
to assured that filter is properly attached.
Note
• The mask can be used without filter.
• Ensure the mask is folded when changing filter.
Discard filter
Pocket Mask is a trademark of
Laerdal Medical AS
© 2017 Laerdal Medical AS.
All rights reserved.
Manufactured in China for:
Laerdal Medical AS
P.O. Box 377, Tanke Svilandsgate 30,
4002 Stavanger, Norway
T: (+47) 51511700
Distributed in US by:
Laerdal Medical Corporation
167 Myers Corners Road,
Wappingers Falls, NY 12590, USA
Tel: +1 (800) 431-1055, +1 (845) 297-7770
FAX +1 (800) 227-1143
Distributed in Canada by:
Laerdal Medical Canada LTD,
151 Nashdene Rd., Unit #45
Toronto, ON, Canada, M1V 4C3
T: +1 (416) 298-9600,
Toll free 888/LAERDAL (523-7325)
ou en français (800) 567-9987
Fax: +1 (416) 298-8016
Printed in China
Storage
Technical Specifications
Changing Filter
Extra oxygen - Connect Oxygen
Tube to Mask Oxygen Inlet.
Oxygen Chart
Material Chart
Oxygen concentration
Inhalation by spontaneously breathing patient
Ventilation of non-breathing patient
Temperature limits
Operating -10 °C (14 °F) to 40 °C (104 °F)
Storage
-40 °C (-40 °F) to 50 °C (122 °F)
Expiratory resistance for the patient*
Mask & One-way Valve
2.9 cm H2O
Mask, Filter & One-way Valve 5.2 cm H2O
Inspiratory resistance for the patient*
Mask & One-way Valve
2.4 cm H2O
Mask, Filter & One-way Valve 4.7 cm H2O
Performance data are nominal values only.
* As per standard test method
Models
This product is in compliance with the essential
requirements of EU Council Directive 93/42/
EEC as amended by EU Council Directive
2007/47/EC.
82 xx xx
Pocket Mask
w/One-way
Valve and Filter
83 xx xx
Pocket Mask w/
One-way Valve,
Filter, Oxygen inlet
and Head strap
Parts
82 06 10
Filter
82 04 10
One-way Valve
Mask in container
Dimensions 13 cm x 11 cm x 4 cm
(5.1 in x 4.3 in x 1.6 in)
Weight
110 g (4 oz)
6310 Rev L
Pocket Mask
User Guide
Place Head Strap
Strapping Mask to face facilitates CPR and
spontaneous inhalation of oxygen.
Select oxygen flow (see oxygen chart).
Provide Mouth to Mask ventilation
if necessary.