19
Maximum position change*
(default 300): Sets maximum range used to search for focus point.
One 1 count = 0.15 um
3.3
Illumination
The LED illuminator channels strobe during each image frame. The Illuminator control is used to
adjust the strobe duration of each channel, effectively altering the LED intensity.
The channels can be adjusted using the controls along the top of the main window, or by
pulling up the LED Pulse Durations window (SVM>LED Settings...).
uScope Illuminator controls
LED Pulse Durations dialog box
When an EPI optics module is installed the forth slider controls the EPI illumination and channel
D is controlled with the Channel B slider/button.
The maximum time duration that can be entered for the pulse duration is dependent on the
camera frame rate, and is equal to 1/f. Checking the Gang box will cause all LED banks to be
adjusted simultaneously when one slider is activated.
Note
:
The Illuminator scroll bars do not update when the illumination is changed via the SVM
front panel. If the front panel will be used for illuminator control, first set all illuminator
slider bars to the maximum setting to allow the full range of illuminator control.