6.4 Hardware Test
In the Hardware test program, you can test if the flowsensor measures the
correct voided volume. Choose
Utilities > Hardware test Bluetooth
to start
the program. Click on the Flowmeter module button to display the following
window.
Figure 6.7
Hardware test Flowmeter
Packet info and Sample rate
Click the
Packet info
button to display module information about the received
and missed packets, and the selected and effective sample rate.
The procedure to test the flowmeter is as follows:
When nothing is placed on the flowsensor, the readout of the field “Weight
absolute” on the screen should be less than 500 grams. If this is not the
case, contact your LABORIE representative.