Hardware
NXT Go Owner’s Manual, LBL-00145, MAN2030
106
Urocap NXT
—Charging the Battery and Pairing
Charge the Urocap NXT by using the supplied wall charger and the NXT Magnetic Pogo Pin
Cable.
Attach the binder connector to the connector on the wall charger.
Connect the wall charger to a power supply.
Attach the Magnetic Pogo Pin connector to the Urocap NXT.
Pair the device by using the USB cable adapter and the pogo pin cable to connect the device
to the laptop PC. Once the device displays in the Synergy
Device Manager
,
disconnect the
cable from the laptop and the laptop PC.
The table below indicates the status of the Urocap NXT and its battery.
Urocap NXT Status
Urocap NXT
Green: Ready to use
Yellow: Not ready to use, commence charging
Lack of Illumination: Device inactive or in sleep
mode; not ready to use.
Device will go into sleep mode one hour after
system shutdown. Return the Urocap NXT to the
charger or use the USB cable to pair to the PC.
Table 13: Urocap NXT LED Signals
LABORIE recommends for the user to have at least 1-2 hours of battery charge on the
Urocap NXT prior to the start of a test or to complete the test with the Urocap NXT plugged
in using the NXT Magnetic Pogo Pin Cable. Always return the device to the Urocap NXT
Charging Hub at the completion of working hours to ensure full charge for next study
period.
Urocap NXT Charging Guidance:
The device battery shall charge from “low” to “charged” in less than 5 hours when in charge
only mode or within 8 hours if charging while acquiring data. “Charged” is battery capacity
sufficient for more than or equal to one hour of recording and “low” is battery capacity not
sufficient for one hour of recording.
Status DRAFT Effective