BA_078_en_10.01 Level transmitter PASCAL Ci4 LEVEL
Page 33/36
6.5.10.1 Device ID
Using the device ID, you can show a custom text in the display if you configure the display
mode accordingly (see 6.3.4). For instance, you can show the tag number continuously in
the display.
The device ID can be up to 16 characters long and consist of numbers, empty spaces,
capital letters and special characters.
For the selection, the characters are arranged in the order shown below. From the end of
the list (special character "@"), you are automatically guided back to the start (number
"0").
Character set:
0123456789 ABCDEFGHIJKLMNOPQRSTUVWXYZ[\
]^_!“#$%&‘()*+,-./:;<=>?@
6.5.10.2 Factory reset
When carrying out a factory reset, all parameters are set to the state at the time of deliv-
ery. This also includes the pressure and current adjustment. Exceptions are the following
operational parameters: “Min/max values”, “HART change counter” and “operation hours”.
6.5.10.3 Configuration memory
You can store all device parameters in the configuration memory of the display module.
You can transfer the data back to the device or to another device. Transferring a configu-
ration to another device requires a compatible target device. The same nominal range is
not necessary. It is sufficient if the measuring range of the stored configuration is equal or
smaller than the nominal range of the target device.
The configuration memory contains the parameter setting of the device at the time of stor-
age. If you change a device parameter afterwards, the stored configuration will not be up-
dated automatically.
The status screen of the configuration memory contains the following information:
- Available
YES: the display module contains a stored configuration
NO: no stored configuration
- Source
this device: the stored configuration originates from the connected device
Serial#: the stored configuration originates from the device with the listed
serial number.
- Up-to-date
YES: The device configuration has not been changed since the parameter
set has been stored in the display module
NO: The device configuration has been changed
The transfer of a configuration takes approx. 25 seconds. If you use a table function with
many support points the transfer can take up to 50 seconds.