Drain the oil mist filter if the oil level is more than 10 mm.
Drain, clean and eventually change the filter insert if smoke comes from the oil mist filter.
9.3
Cleaning
NOTE
When using strong materials such as strong acids or bases, radioactive substances
and volatile organic solvents it must be handled carefully and promptly cleaned up if
spilled.
If a sample is spilled in the concentrator chamber, it must be cleaned up immediately.
Cleaning of acrylic components use only mild detergents solutions NEVER solvents.
Wipe stainless steel surfaces with alcohol and then dry, do not use abrasives.
The condenser of the CoolSafe once drained, wipe dry and then lightly cleaned with alcohol,
ensure that the drain valve is open. Do not forget to close the valve once the condenser
has dried.
9.4
Checking for Leaks on the CoolSafe freezer dryer
Always start from the vacuum pump and work forwards towards the chamber.
1.
Check vacuum hose connection to the pump. If the pump is fitted with a nozzle and if this is
held in place with a clamp, check that there is a centring ring with seal fitted between the
flange on the pump and the nozzle. Make sure the clamp is tight and that the hose is pushed
well onto the nozzle, and that the clip on the hose is tight.
2.
Check the other end of the vacuum hose where it connects to the CoolSafe, make sure the
clip is tight.
3.
The drain valve on the front of the Cool Safe also has to be checked for leaks. First unscrew
the drain valve (with black knob) all the way until its loose and then pull towards you. The
barrel of the valve then comes away from the body. On the end of the barrel there is a white
Teflon plug, make sure this is clean and without any scratches, also wipe the inside of the
valve body. Replace the valve barrel into the body of the valve, push in and then screw the
valve completely to the fully closed position. When vacuum is applied check that there is no
leak from the drain-this can be done placing a finger over the drain spigot, you will feel if
there is a leak.
4.
Check the seal on the acrylic plate of the chamber that interfaces with the top stainless steel
part of the CoolSafe. This can be done by squirting or brushing some frothy soap solution
all around, if there is a leak this will be sucked in towards the condenser.
5.
If leak(s) are found on the seals, clean both the surface of the s/steel and the seal, using a
clean tissue slightly wetted with alcohol.
6.
Check that the lid of the chamber is located correctly, again check that the seal is OK as
previous.