 1.0  Introduction

The Labnet Orbit™ Digital Shakers are intended for shaking microliter plates, tubes, bottles, flasks, dishes and other laboratory 
vessels. The shakers are driven by asynchronous motors, which enable silent operation and constant shaking speed independent of 
the load or power supply fluctuations.

Each shaker device consists of two main components:

 



Motor with eccentric mechanism

 



Control electronics

The motor drives the shaker’s eccentric mechanics and generates shaking effect. The electronics control the motor RPM, TIME, and 
keyboard functions. 

 2.0  Safety  Precautions

 



Always plug the unit into a properly grounded and fused outlet.

 



Do not use the device near water sources. Take care so that water will not fall on the device, especially during cleaning procedures.

 



Do not use the device in a caustic or explosive atmosphere.

 



There are no user serviceable parts in the unit. Opening the unit may void the warranty. In case of a malfunction or liquid being 
spilled into the unit, unplug the device from its power outlet and contact Corning Customer Service.

NOTE: 

In the case that the device does not function properly even if you have exactly followed the instructions described in 

the User’s Manual, contact Corning Customer Service. If the equipment is used in a manner not specified by this manual, the 
equipment may become unsafe to operate, could harm the user or the device, and may void the warranty.

Do not shake inflammable or explosive samples or use the unit in an inflammable or explosive environment.

 3.0  Specifications

The casings of the Labnet Orbit P4, 300, and 1000 are made of plastic. Bottom cover and upper plate are made of steel plate varnished 
with high resistant polyurethane lacquer.

The housing of the Orbit 1900 is made from a high-grade cold rolled steel plate and painted with a highly resistant polyurethane 
lacquer.

 



Labnet Orbit P4

Electric Power Supply

120V 
230V

120V ± 10%, 50/60Hz 
230V ± 10%, 50/60Hz

Motor Power

15W

Fuse

120V 
230V

2 x 1AT, 120V 
2 x 0.25AT, 230V

RPM Regulation

Digital, load independent, from 100 to 1,200 rpm (1,400 rpm, 120V version) in 10 rpm steps

Shaker Orbit

3 mm

Timer

30 sec. to 99 min. 50 sec. in 10-second steps (under 10 min. in 1-second steps), timer HOLD function

Maximum Load

0.66 lbs. (0.3 kg)

Dimensions (W x L x H)

7.4 x 11.8 x 5.9 in. (18.8 x 30 x 15 cm)

Weight

9.5 lbs. (4.3 kg)

Ambient Operation Range

5°C to 40°C, up to 85% RH, non-condensing

Summary of Contents for Orbit S2020-P4-B

Page 1: ...Labnet Orbit Digital Shakers Instruction Manual Catalog Numbers S2020 P4 B S2020 P4 B 230V S2030 300 B S2030 300 B 230V S2030 1000 B S2030 1000 B 230V S2040 1900 S2040 1900 230V...

Page 2: ...3 4 2 Selecting the Location 3 4 3 Attaching the Power Cord 3 4 4 Installing Shakers Platforms 3 5 0 Operating Instructions 4 5 1 Control Panel 4 5 2 Basic Operation 5 5 3 Additional Operations 6 6 0...

Page 3: ...that the device does not function properly even if you have exactly followed the instructions described in the User s Manual contact Corning Customer Service If the equipment is used in a manner not...

Page 4: ...25AT 230V RPM Regulation Digital load independent from 20 to 300 rpm in 1 rpm steps Shaker Orbit 19 mm Timer 30 sec to 99 min 50 sec in 10 second steps under 10 min in 1 second steps timer HOLD functi...

Page 5: ...the following are correct Model Serial number Electrical rating Should any kind of damage have occurred during transportation contact the carrier immediately The carrier is responsible for shipping d...

Page 6: ...with new stabilizers and set to appropriate height Labnet Orbit 1000 and Orbit 1900 The Orbit 1000 and Orbit 1900 shakers have a variety of platforms to meet most shaking needs These platforms mount...

Page 7: ...ting the encoder knob right or left the set time can be selected from 30 sec to 99 min 50 sec 99 5 99 min 50 sec 9 59 9 min 59 sec 0 30 30 sec If you want to set the timer to Hold turn the knob left o...

Page 8: ...after 2 sec When the run time has elapsed or when you press the START STOP button End will show on the display and the Run light will flash When the shaker completely stops the set speed and time will...

Page 9: ...it require return to Corning for service it should be properly packed to avoid damage Any damage resulting from improper packaging shall be the responsibility of the user 7 0 Maintenance and Cleaning...

Page 10: ...y Pk S2040 12 Flat platform with non slip mat 40 x 50 cm 1 S2040 15 Universal bar clamp platform for holding a variety of containers 1 S2040 18 Flask clamp platform clamps sold separately 1 Clamps for...

Page 11: ...ces or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products...

Page 12: ...IN AMERICA grupoLA corning com Brazil t 55 11 3089 7400 Mexico t 52 81 8158 8400 To request certificates please contact us at www labnetlink com Warranty Disclaimer Unless otherwise specified all prod...

Reviews: