Introduction
1-17
1.3.5 1536-plate to 1536-plate liquid transfer
Fluid transfers between two 1536-well microplates are exactly the same as
384-plate to 384-plate transfers, except for the well densities.
Note:
Echo 555 only: High-throughput transfer of DMSO is available
for the 1536-well LDV plate (1536LDV_DMSO_HT). This special
calibration type uses the same 1536-well LDV plate that is used for
standard DMSO and aqueous transfer; however, the repetition rate for
droplet transfer is much faster. This calibration type is ideal for
preparing multiple screening plates and prewetted plates.
Note:
For liquid transfer of very small volumes, or transfer that
requires precise droplet placement, use the 1536LDV_DMSO
calibration.