Echo Liquid Handler Software
5-47
5.4.2 Power calibration
Proper acoustic droplet ejection relies on a calibrated RF energy source.
Environmental temperature changes can affect the characteristics of the
Echo RF energy pulses. Power calibration is performed automatically by the
Echo liquid handler as part of each liquid transfer protocol.
Note:
Normally a manual power calibration will not need to be
performed by a user. However, it can be useful during certain repair,
diagnostic, or recovery procedures. Perform power calibration only as
directed by Labcyte personnel.
If manual power calibration is required, use the following procedure:
1.
On the
C
ALIBRATION
tab, click the
C
ALIBRATE
button in the
P
OWER
section.
The
C
ALIBRATE
P
OWER
dialog box will open.
Figure 5.44 Calibrate Power dialog
box reporting current values
Reference
Amp
: The
A
MP
field displays the reference tone burst amplitude.
Pulse Energy
: The
P
ULSE
E
NERGY
field displays the reference pulse
energy value.