Prudence:
Le gaz naturel devrait être seulement utilisé dans l'accessoire de service qui
a été pré soudé avec des tuyaux de cuivre. Le contenu soufré du gaz pourrait causer la
détérioration des lignes d’alimentation en cuivre standard.
Step 4
When the filtered hood has been set in place, wired, and plumbed, it should be sealed
at the work surface to prevent spilled materials from collecting under the walls of the
filtered hood. Materials such as silicone sealants are recommended to seal the hood
structure.
Filter Installation
This section will detail how to install and check the filtration for the hood, including
detailed figures and pictures (all figures will be shown for M3 or 5FT filtered hood). If
using the HEPA filter option, reference
Section 8: Maintaining your filtered hood
HEPA filter installation and replacement
. This section describes how to:
Confirm order and check supplies.
Locate the power supply, command module and anemometer
Unpacking pre-filters, filters and fan boxes
Place pre-filters and stack on primary filters
Stack fan boxes on primary filters
Place secondary filters on fan box
Label all filters and fan boxes
Add protective caps to secondary filters
Step 1
Collect the following tools required to install the equipment:
6ft ladder
Tape and a marker
2 people required
Step 2
To confirm your order, make sure you have the correct number of fans for the filtered
hood size:
M1 (1 column filtration) 3ft Protector Airo
M2 (2 column filtration) 4ft Protector Echo/Airo
M3 (3 column filtration) 5ft Protector Echo
M4 (4 column filtration) 6ft Protector Echo
M5 (5 column filtration) 8ft Protector Echo
Step 3
Each Protector Hood will have the following Neutrodine filters based on their size per
each filtered hood installed.
M1 (1 column filtration) 2 total Neutrodine filters
Summary of Contents for Protector Airo
Page 78: ...78 Fans and Lights not working ...
Page 79: ...79 Vertical sash no longer operates smoothly ...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working ...
Page 81: ...81 Contaminates outside the filtered hood ...
Page 82: ...82 Lights operate but fans will not ...
Page 83: ...83 Smart Command Alarms ...
Page 84: ...84 If needed contact Labconco to troubleshoot further ...