18
Carbon Filter Life
It is
very
important to change the Main Carbon Filter in the CApture BT Fuming
Chamber when prompted (see
Section 7: Replace Main Carbon Filter
) or sooner if
Cyanoacrylate (CA) fumes are detected emitting from the chamber
’s exhaust.
Never
unseal the protective bag around a Carbon Filter before it is ready to be installed in the
unit. Review the chart below to determine the appropriate number of Filter Life Cycles
for your unit.
On the Main Menu, Select
SERVICE
, press
[OK]
. The password is:
[UP] [DOWN]
[LEFT] [RIGHT] [OK]
.
Maintain password only with responsible party. Press buttons in
proper order to gain access to the
SERVICE
menu. After entering the password, select
LIFE CYCLE
, and press
[OK]
. The Life Cycle menu (Fig. 3-1) is displayed:
Using the UP/DOWN Buttons, adjust the Filter Life Cycle number as desired between
200 and 500 Cycles in 50 Cycle increments. Press
‘OK’
to accept.
1-2
3-4
5+
<1.5
500
400
350
1.5-2.5
400
350
300
2.5+
350
300
250
Avg. Cycles per Day
A
vg
.
Gra
m
s
CA
p
e
r C
ycle
•
Not regularly cleaning CA deposits
on the inside of the chamber
•
Not replacing the Pre-Filter regularly
•
Enabling the Standby Purge feature,
which helps remove CA fumes
Additional factors that will
reduce Filter Life:
Figure 3-1
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...