Changing the Optional UV Lamp
The UV Lamp is an optional feature and may not be found on all models.
For optimum performance, the UV lamp should be changed as indicated by the UV
lamp hourmeter.
The UV lamp and the work area of the cabinet must be thoroughly decontaminated
before removing the lamp.
1. Start the cabinet blower and let it operate for 5 minutes.
2. Raise the sash to its full open position.
3. Thoroughly surface decontaminate the UV lamp and the work area of the cabinet.
4. Unplug the cabinet or turn off the System Reset Switch, located on the top of the
cabinet.
5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and out of its
sockets.
6. Install new lamp by reversing the removal procedure.
THE UV LAMP IN THIS PRODUCT CONTAINS MERCURY
Manage in accordance with local disposal laws. DO NOT place lamps in trash.
Dispose as a hazardous waste. For information regarding safe handling,
recycling and disposal, consult www.lamprecycle.org
CETTE LAMPE (UV) DANS CE PRODUIT CONTIENT DU MERCUE
Éliminez ou recyclez conformément aux lois applicables. Pour de l‘information
concernant des pratiques de manipulation sécuritaires et l’élimination sécuritaire
et le recyclage, veuillez consulter www.lamprecycle.org
Summary of Contents for Axiom C1
Page 2: ...Axiom Type C1 Biosafety Cabinets 2021 Present 30441xxx1 30448xxx1 30461xxx1 30468xxx1...
Page 75: ...75 Blower only will not start...
Page 76: ...76 Lights only will not illuminate...
Page 77: ...77 UV Light will not illuminate...
Page 78: ...78 Airflow Alert activating...
Page 79: ...79 Filter Life Gauge not at 100 when new...
Page 80: ...80 Contamination in the work area...
Page 82: ...82 Figure A 1 1 2 3 4...