142
12. APPENDIX
12-1 LIST OF SETTING ITEM
12-1-1. Calibration
Item
Function Name
Parameter Name
Setting Value
Initial Value
Manual Input Cal. Bridge Voltage
0:2 V, 1:10 V
2 V
No-Load Zero
(-3.2000 to 3.2000 mV/V)
(0.000 mV/V)
Sens. Regist. Cal.
Rated output
-3.2000 to 3.2000 mV/V
2.0000 mV/V
Rated display
-99999 to 99999
10000
Actual Load Cal.
(Rated output)
(-3.2000 to 3.2000 mV/V)
(2.000 mV/V)
Rated display
-99999 to 99999
10000
Unit
79 types
(Refer to “12-2 UNIT LIST.”)
None
TEDS Auto Cal.
TEDS Part Cal.
TEDS Cal. Item
Applied voltage
Execute, prohibit
Execute
Rated output
Execute, prohibit
Execute
Rated capacity
Execute, prohibit
Execute
Unit
Execute, prohibit
Execute
TEDS Ope..
Setting
TEDS Reading Ope.
Key operation
Allow, prohibit
Allow
Control input
Allow, prohibit
Allow
Power ON
Allow, prohibit
Prohibit
Command
Allow, prohibit
Allow
Zero during TEDS
Execute, prohibit
Execute
TEDS Info Display
Numeric Regist
Cal.
Initial Value
-3.2000
~
3.2000 mV/V
(0.000 mV/V)
12-1-2. Input Setting
Item
Function Name
Parameter Name
Setting Value
Initial Value
Smoothing
Analog Filter
0: 1 Hz, 1: 30 Hz, 2: 300 Hz,
3: None (1 kHz or more)
1Hz
Minimum Scale
0: 1, 1: 2, 2: 5, 3: 10
4: 20, 5: 50, 6: 100
1
Moving Average
00: None, 01: 2 times,
02: 4 times,03: 8 times,
04: 16 times, 05: 32 times,
06: 64 times, 07: 128 times,
08: 256 times, 09: 512 times,
10: 1024 times, 11: 2048 times
None
Zero
compensation
Zero tracking
Judging time
0.00 to 9.99 sec.
0.00 sec.
Compensation range
-99999 to 99999
0
Zero nearly zero
0 to 9
0
Additional Value
-99999 to 99999
0
Original Value
(-3.2000 to 3.2000 mV/V)