126
(28) [SMF]
Content:
Set the measuring mode setting.
Format:
SMFaa, b, c, x. xx, y. yy<CR>
Parameter:
aa, b, c, x. xx, y. yy
Detect time: 0.00 to 9.99 (sec.)
Delay time: 0.00 to 9.99 (sec)
Comparison mode
0: Normal Comp.
1: Hold Comp.
Display mode
0: Normal Disp.
1: Hold Disp.
Operation mode
00: Normal
01: Peak hold
02: Block-specified peak
03: Time-specified peak
04: Bottom hold
05: Block-specified bottom
06: Time-specified bottom
07: Arbitrary point hold
08: Block Peak-Bottom
09: Time Peak-Bottom
10: Block average
11: Time average
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR070
Return value
aa, b, c, x.xx, y.yy<CR><LF>
Detect time
Delay time
Comparison mode
Display mode
Operation mode
(29) [SG1]
Content:
Set the waveform display setting 1 (X-axis and Y-axis settings)
Format:
SG1a, b, ±xxxxxx<CR><LF>
Parameter:
a, b, ±xxxxxx
Start point of Y-axis: -99999 to 99999 (count)
Indicated value with a decimal point. If the indicated value has no
decimal point, it is filled with ‘0s’ from the head.
End point of X- axis
0: 0.5 sec.
1: 1.0 sec
2: 2.0 sec
3: 5.0 sec
4: 10.0 sec.
5: 20.0 sec.
6: 50.0 sec. 7: 100.0 sec.
End point of Y-axis
0: 250
1: 500
2: 1000
3: 2000
4: 5000
5: 10000
6: 20000
7: 50000
8: 100000
9: 200000
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR130
Return value
a, b, ±xxxxxx<CR><LF>
Start point of the Y-axis: -99999 to 99999 (count)
End point of the X-axis
End point of the Y-axis