(24) [SZM]
Content:
Set the additional value.
Format:
SZM1, ±xxxxxx<CR>
Parameter:
±xxxxx
-99999 to 99999 (count)
Indicated value with a decimal point. If the indicated value has no
decimal point, it is filled with ‘0’ from the head.
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR030
Return value
±xxxxxx<CR><LF>
Additional value
(25) [SZT]
Content:
To set Automatic ZERO compensation functions
Format:
SZTxxxxxx, y.yy, z<CR>
Parameter:
xxxxxx, y.yy, z
Zero-near-zero: 0 to 9 (count)
Determination time: 0.00 to 9.99 (sec)
Automatic ZERO compensation range: 0 to 99999 (count)
The value, including the decimal point. When the value has no
decimal point, it is filled with “0” from the head.
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR090
Return value
xxxxxx, y.yy, z
Zero-near-zero.
Determination time.
Automatic ZERO compensation range.