Chapter 17 - Management of the instrument 129
10713 Vxx.11
17.11
Configuration of PING test in the DOCSIS analyzer
You can configure a Ping test by selecting
MODE
->
Settings
->
DOCSIS analyzer
->
Ping Test
.
64 byte long PING packets are sent as standard during the test. The number of packets to be sent
and the time interval between two PINGS can be configured in this submenu.
You can set the number of packets to be sent by selecting
MODE
->
Settings
->
DOCSIS analyzer
->
Ping Test
->
Packets
. The number of packets is set after opening the submenu. You can use
the numerical keypad to change this number. The permissible value range is 5 to 100 packets. If a
value below 5 or above 100 is entered, the display will switch to the last correctly entered value.
After entering a valid value and pressing
ENTER
twice, the setting is applied for all future PING
tests until a new value is entered. The entry is non-volatile and is saved even when the instrument
is switched off.
Select
MODE
->
Settings
->
DOCSIS analyzer
->
Ping Test
->
Interval
to set the interval
between two PINGs. You can choose between 0.5, 1, 2 and 5 seconds. Exit the submenu by
pressing
ENTER
. The entry is non-volatile and is saved even when the instrument is switched off.
17.12
Deactivating the DOCSIS analyzer
The DOCSIS analyzer can be deactivated manually. This can be useful in many cases.
If no return-path-capable components are integrated in a system, but the DOCSIS downstream
channel is still to be measured, this will prevent the measuring instrument on the upstream from
trying to send. As this occurs in such a case with a high transmitting power, the downstream
measurements can be interrupted in some circumstances.
With automatic measurements (DataLogger), the upstream transmitting power is also taken into
consideration with the active DOCSIS analyzer. However, to do this, the ranging process must be
completed. This usually takes a few seconds. If only the downstream parameters are to be
recorded with DOCSIS, it is better to deactivate the DOCSIS analyzer.
MODE
->
Settings
->
DOCSIS analyzer
->
Upstream
can be used to activate and deactivate the
DOCSIS analyzer. The menu item for the currently active setting (on or off) is displayed in an
inverted manner.
This setting is non-volatile. It is also incorporated in the tuning memory. For this reason, memory
locations can be created with an active or inactive DOCSIS analyzer. Memory entries, for which the
DOCSIS analyzer is active are marked with
+US
.
In the default setting, the DOCSIS analyzer is active.