Kurz
B
‐
Series
Operations
Guide
6–
2
Flow Meter Configuration
Specifying Drift Check Triggers
There
are
four
ways
to
initiate
the
zero
‐
mid
‐
span
drift
check:
•
A
specified
time
interval
•
An
external
contact
closure
through
digital
input
#2
•
A
menu
option
in
Extended
Utilities
mode
•
A
Modbus
command
The
duration
of
the
individual
zero,
mid,
or
span
checks
are
controlled
by
user
‐
specified
duration
values.
Specified-Time Interval Drift Check
The
zero
‐
mid
‐
span
check
can
be
triggered
by
the
drift
check
interval
time.
To
configure
this
option,
turn
ON
the
Auto
Drift
Check.
When
it
is
triggered,
it
will
perform
the
zero,
mid,
and
span
sequence.
Change
the
value
of
the
drift
check
interval
to
change
the
frequency
that
the
drift
check
starts.
See
for
information
about
setting
up
drift
check.
External Digital Input
The
external
digital
input
trigger
does
not
require
that
Auto
Drift
Check
be
turned
ON.
When
the
flow
meter
digital
input
(DI)
channel
#2
is
grounded,
the
meter
initiates
the
zero
‐
mid
‐
span
drift
check
sequence.
While
the
system
is
busy
performing
the
drift
check
sequence,
it
will
not
accept
another
command
via
the
digital
input.
After
the
drift
check
sequence
completes,
the
meter
monitors
the
DI
channel
for
the
next
request
to
initiate
a
drift
check.
If
a
specified
time
internal
drift
check
is
underway
while
an
external
drift
check
is
triggered,
the
external
drift
check
will
start
after
the
internal
drift
check
completes.
Summary of Contents for 410FTB
Page 1: ...Operations Guide for B Series Flow Meters 368042E...
Page 4: ...Kurz B Series Operations Guide iv...
Page 9: ...Kurz B Series Operations Guide ix Preface...
Page 33: ...Kurz B Series Operations Guide 1 20 B Series Flow Meter Modes Menus...
Page 65: ...Kurz B Series Operations Guide 2 32 B Series Flow Meter Display Mode Options...
Page 141: ...Kurz B Series Operations Guide 4 68 Program Mode Advanced Setup...