Program Mode
Advanced Setup
Changing to Another Gas Curve Manually
Changing
to
another
calibration
curve
using
Manual
mode:
1>
Press
P
.
2>
Enter
the
Advanced
access
code,
and
then
press
E
.
3>
Press
2
to
invoke
the
Quick
Jump
option.
4>
Press
17
for
the
Calibration
Curve
Selection
menu,
and
then
press
E
.
The
prompt
for
configuring
the
source
of
the
gas
calibration
curve
appears.
CURVE SEL MODE
>MANUAL ^v
5>
Use
the
arrow
keys
to
select
MANUAL
,
and
then
press
E
.
The
prompt
for
the
first
gas
curve
appears.
SELECT CAL CURVE
>50CH4+40CO2+ ^v
Each
curve
can
be
a
single
gas
or
a
gas
mix.
For
gas
mixes,
the
entire
gas
mix
name
might
extend
beyond
the
available
character
spaces
in
the
display.
Refer
to
your
calibration
data
sheet
for
complete
information.
6>
Press
the
up
or
down
arrow
to
scroll
through
the
list
of
available
gas
curves.
7>
Press
E
to
select
a
specific
gas
curve.
MOLWT 28.997
KG/M3= 1.157254
The
molecular
weight
and
calculated
density
of
the
gas
mix
appears.
8>
Press
E
to
return
to
the
main
menu.
Summary of Contents for 410FTB
Page 1: ...Operations Guide for B Series Flow Meters 368042E...
Page 4: ...Kurz B Series Operations Guide iv...
Page 9: ...Kurz B Series Operations Guide ix Preface...
Page 33: ...Kurz B Series Operations Guide 1 20 B Series Flow Meter Modes Menus...
Page 65: ...Kurz B Series Operations Guide 2 32 B Series Flow Meter Display Mode Options...
Page 141: ...Kurz B Series Operations Guide 4 68 Program Mode Advanced Setup...