QuickGene-810 Operation Manual
Rev. 1.7
13
1 Installation
1
3
2
5
4
6
7
■
Setup the Tube Cap
The following explains how to install the Tube Cap.
(1) Set the Tube Cap on the conical tube.
•
Set the purple cap on the conical tube for
wash buffer (right).
•
Set the blue cap on the conical tube for elution
buffer (left).
(2) Connect the Tube Cap and the Pump
Tube.
Check the combination of the Pump Tube and Tube
Cap label, and then turn the connector to connect
them as shown in the figure.
•
Connect the Pump Tube for the Wash Buffer
Pump (upper, purple) and the purple Tube
Cap.
•
Connect the Pump Tube for the Elution Buffer
Pump (lower, blue) and the blue Tube Cap.
■
Setup the Holder Carriage
The following explains how to install the Holder Carriage.
(1) Attach the Holder Stand to the
QuickGene-810.
Insert the tabs of the Holder Stand in the slits on the
QuickGene-810.
(2) Set Tube Holder and Cartridge Holder in
the Holder Carriage.
Tube Cap (Blue)
Tube Cap (Purple)
Tube Cap (Blue)
Tube Cap (Purple)
To Wash Buffer Pump
To Elution Buffer Pump
Note: After attaching the Holder Stand,
check that there is no gap between
the bottom of Holder Stand and the
QuickGene-810.
PINCHING OF
FINGERS:
Do not insert fingers under
the table unit.
Holder Stand
Table unit
Holder Carriage
Tube Holder
Cartridge Holder
Tab
Align the tab
and notch
Summary of Contents for QuickGene-810
Page 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Page 2: ......
Page 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Page 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Page 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Page 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Page 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Page 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Page 76: ......