Krypton Neon DIS Installation and Set Up Guide 

 

Kuntze Instruments 

 

 

 

 

 

 

3316 Leechburg Road | Lower Burrell | Pennsylvania | 15068 
724-339-8510 | [email protected] | kuntzeinsturments.com 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

General Steps 

1.  Mount the system upright on the wall 

2. 

Insert clean sensor 

Insert Disinfection sensor into the Flow Cell Chamber for the 
Disinfection Sensor (5) and screw on the Disinfection Cable 

3.  Power on system 

4. 

Check flow 

5. 

Calibrate sensor 

6.  If needed, set user specific settings 

– see operations manual 

a.  ASR 

b.  4 

– 20 mA 

c.  Relay Alarms  

 

 

 

 

Key 

1

 

– Neon DIS Instrument 

2

 

– Argon Stabiflow  

3

 

– FTG Temperature Cable 

4

 - Disinfection Cable 

5

 

– Flow Cell Chamber for Disinfection Sensor 

6

 

– Water Inflow Valve 

7

 

– Water Outflow Valve 

8

 

– Water Sampling Reference Valve 

9

 

– Mounting Board 

Check Your Flow 

Make sure the Water Sampling Reference Valve is in the closed position.  

System needs proper 

flow to operate. “Digital input 1 – no water” will show on display screen if 

flow is not strong enough. 

Flow minimum: 5 psi 

 

Flow maximum: 87 psi 

 

Power Requirements 

System comes with US power cord 

Line Voltage

85-265 V AC, 50-60 Hz 

Power Consumption

10 VA 

 

 

 

 

Cleaning Sensor 

IMPORTANT

: Sensor must be cleaned by 

hand before being inserted into the chamber!  

How to clean DIS sensor: 

- Remove sensor from assembly and cable 
- Wipe off any pollution with paper tissue, rinse     
   with water 
- Clean metal bands with powder detergent (ex.   
  Comet), wipe gently 
- Rinse with water and reinstall 

 
 
Calibrating Sensor 

DIS Sensor 

1. 

IMPORTANT: Clean sensor before 
proceeding with calibration! 

2.  Using water sample taken from Sampling 

Reference Valve, perform DPD test 

3.  Menu > Cal > Cal DIS > enter DPD 

reference value > Ok > Save

 

 

 