INSTALLATION
3
13
OPTIFLUX 5000
www.krohne.com
03/2017 - 4000686204 - MA OPTIFLUX 5000 SW R04 en
á
Small screwdriver
á
Wrench for cable glands
á
Wrench for wall mounting bracket (remote version only)
á
Torque wrench for installing flowmeter in pipeline
3.5 General requirements
3.5.1 Vibration
3.5.2 Magnetic field
INFORMATION!
The following precautions must be taken to ensure reliable installation.
á
Make sure that there is adequate space to the sides.
á
Protect the signal converter from direct sunlight and install a sun shade if necessary.
á
Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.
á
Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-64.
Figure 3-2: Avoid vibrations
Figure 3-3: Avoid magnetic fields
3.4 Pre-installation requirements
Make sure that you have all necessary tools available:
á
Allen key (4 mm)