5.3Power Off

In normal detection mode long press

power button for five seconds until the “DI” sound ,display shows shutting

down as it show in figure 5,device will be shut off within 1 minute .

6.Operation Interface

6.1 Gas Detection Interface

Under normal-detection mode,according to the number
and types of sensors in the device,there are different
interfaces: single gas detection mode as shown in
figure6(O3 for example),two kinds of gas detection mode
as shown in figure 7 (O3

H2S for example) ,three kinds

of gas detection mode as shown in figure 8(O3

H2S

CO

for example),four kinds of gas detection mode as shown
in figure 9(O3

CO

H2S

CH4for example).

Icons at the top of the display shows the “pump status”,”battery capacity”,”humidity”,take figure 9 for example,there are four channels,
figure indicate corresponding gas concentration in each channel,”O3” is the gas’s molecular formula,lower right”PPM” is the
concentration unit;when one or several gas channel reach the alarm value,there will be a alarm sign under corresponding

Summary of Contents for GT-1000 SERIES

Page 1: ...ShenZhen Korno Import Export Co Ltd http szkorno en made in china com 1 GT 1000 SERIES MULTIFUNCTIONAL GAS AND DUST DETECTOR USER MANUAL ShenZhen Korno Import Export Co Ltd...

Page 2: ...arm or death Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings Please calibrate and maintain this instrument regularly Frequency of calibra...

Page 3: ...tion Interface 14 6 1 Gas Detection Interface 14 6 2 Function Menu Instruction 15 6 2 1 Basic Setting 15 6 2 2 History Data Log 16 6 2 3 RealTime Curve 17 6 2 4 Zero Calibration 18 6 2 5 Target Calibr...

Page 4: ...es the function of seven buttons Detector interface Menu parameter setting Up Save instant data press and hold for 5 seconds manually saving mode Up Move up Value Down Save instant data press and hold...

Page 5: ...of the detector which might affect the service life and precision of the detector or even directly damage detector When a user accidentally makes out range operation he should evacuate the instrument...

Page 6: ...m had been preset for all GT 1000 multifunctional gas detector when no operation were made within 30 seconds the protection program will activate and the display of the detector will dim out user can...

Page 7: ...r technology GT 1000 series multifunctional gas detector can detect corresponding gas and dust concentration at the same time rapidly and precisely we maintain a leading position in domestic level and...

Page 8: ...le printer to print data With temperature and humidity detection user can detect temperature and humidity value on the scene or the temperature and humidity value inside the pipe Five operation modes...

Page 9: ...is optional Temperature and humidity Temperature detection range 40 120 Humidity detection range 0 100 RH Detection pattern Pumping with built in high power pump allow device working under tiny negat...

Page 10: ...ial ABS PC ABS PC IP rating IP66 Working temperature 30 60 Power 4000mA high capacity polymer rechargeable battery Humidity 95 RH Non condensing Dimensions 220 88 55 mm L W H 0 5 Kg net weight Working...

Page 11: ...ShenZhen Korno Import Export Co Ltd http szkorno en made in china com 11 4 Product Structure...

Page 12: ...ng form is description of the seven buttons Detector interface Menu parameter setting Up Save instant data press and hold for 5 seconds manually saving mode Up Move up Value Down Save instant data pre...

Page 13: ...normally Only in condition of the pump was switched on 5 2Power On Long press power button for five seconds until you hear a beep the DI sound then the display and red signal light turn on The screen...

Page 14: ...ction mode as shown in figure6 O3 for example two kinds of gas detection mode as shown in figure 7 O3 H2S for example three kinds of gas detection mode as shown in figure 8 O3 H2S CO for example four...

Page 15: ...button press Ok to enter corresponding sub menu press Back to return to normal detection interface or previous menu 6 2 1 Basic Setting In Basic setting menu user can see various setting as shown in...

Page 16: ...when the detector is equipped with VOC sensor and dust particle sensor The user can choose the specific VOC gas name as shown in figure18 and choose the detection unit or the min detect diameter of d...

Page 17: ...his menu users can view all the sensors and corresponding data quantity press OK again to history data of specific channel as it shown in figure 27 5 Clear ALL History Log Press OK to delete all histo...

Page 18: ...in fresh air or high purity inert gas for example 99 999 VOL N2 etc 2 Do not operate zero calibration for those gases which already exist in the air such as oxygen carbodioxide nitrogen 6 2 5 Target C...

Page 19: ...erate stand for this level has been calibrated and stand for this level still need to be calibrate first of all input a concentration value of standard gas then calibrate Target gas concentration valu...

Page 20: ...l between the two fixed time measure as shown in figure 37 6 2 8 Store Setting Manually save automatically save are optional in this menu it also allows user to set storage cycle under automatically s...

Page 21: ...o set date and time this time is related to the time of the concentration values are being recorded as it shown in figure 42 6 2 11 Factory Setting If user proceeded an wrong operation by accidentally...

Page 22: ...high you need to proceed zero calibration Problem No response or weak response when detecting Possible reasons 1 Oxygen content value of gas is too low 5 VOL 2 Gas pressure is too high the pump can t...

Page 23: ...or inhaled Solutions Return to manufacturer to replace the pump install a dust and vapor filter at the gas inlet Problem Unable to boot up instrument Possible reasons Battery low or empty Solutions Tr...

Page 24: ...face of the software will show connection port standby defaulted baud rate is 115200 do not change 4 Enter the history data interface to output datalog there will be menu prompts at the both software...

Page 25: ...alue User should prevent drop or impact the detector It is strictly forbidden to use detector in high temperature high humidity or high pressure environment if workingenvironment is high humidity dete...

Page 26: ...m 26 ShenZhen Korno Import Export Co Ltd ADD 4E Building 1 NO 6 LiuFang Road 67District BaoAn ShenZhen TEL 86 0755 86110165 666 FAX 86 0755 86110165 555 Web Site http szkorno en made in china com To g...