15
WARNING
Oxygen uptake and atomization cannot be used at the same time.
Adjust the flow rate to minimum value when using the atomization
function.
3.8
Other operations
3.8.1
Timing operation
During working the concentrator can be set timer length by using timer
function.
1.
Press the I/O power switch to position "
I
" (ON), and then press
to
start the oxygen supply.
2.
Press button,
Timing (
,
-)
to set timer timing.
3.
Using
Increase
(
+
) button to adjust the timing hours (maximum is 10
hours).
4.
Then press
to confirm the setting.
When the shut hour is set, the system comes into count down time and the
oxygen concentrator LCD will shows the remaining time. When remaining
time becomes 0, the oxygen concentrator will go to standby status.
3.8.2
Remote control operation
The function of buttons on the remote controller is the same with some
keys on the concentrator display panel. Refer to the previous chapter
description for the details of buttons function.
Press
on the remote controller when the concentrator power is on and
in standby status.
Press
while the concentrator operating to stop the oxygen supply.
Press timer button on the remote controller to set timer length.
When the remote controller cannot work, the battery may not have
sufficient power. Please replace the battery periodically with the same
battery and be careful of the affirmative and negative pole.
Remaining time