14
Every few seconds the concentrator issues "TAPOO" sound while operating,
it is the normal ventilation sound.
WARNING
Choose safety-qualified socket and socket board with safety electrical
certification.
Do not remove the cabinet unless qualified technician, in case of
damaging the machine, electric shock or other accident.
Oxygen uptake time and oxygen flow adjustment should follow the
physician's advice.
Check that the tubing or accessory is clogged, kinked or the humidifier is
damaged if the flow value is lower than 0.5L/min.
If no use of concentrator for a long-term, the water in humidifier bottle
should be drained. Save the humidifier bottle after cleaned and dried.
3.7
Atomization operation of N, M type oxygen concentrator
(not administered)
1.
Press the power switch to position “
I
” (ON) to turn on the
concentrator.
2.
Connect the atomized mask or mouthpiece to the medicine cup.
3.
Unscrew the nut of the oxygen atomizer outlet; connect the
atomization tubing to the outlet.
4.
The press
to start oxygen supply, and using knob to adjust the flow
to minimum.
5.
Put the atomization mouthpiece into the mouth or wear the mask and
start the atomization therapy (follow the doctor's instructions for the
atomization time).
6.
After the atomization is finished, press
to stop atomizing. Remove
tubing and turn off the oxygen concentrator by pressing the I/O power
switch.
Atomization performance
The atomization outlet pressure is 60kPa to 250kPa, and the maximum air
flow rate is ≥
10L/min.
The atomization rate of the atomization assembly is ≥0.2 mL/min.