In the interests of safety, it is the user’s responsibility to carry out the following checks before and after using the sys-
tem . If the system seems abnormal after checking the items below, contact your service representative .
Keep the maintenance records using the "Check List" .
Refer to "11 .7 Check List for Maintenance Check: Before and after use" .
7 .1 .1 Check before Turning ON the Power
Check the items below before turning ON the system .
No .
Item
(1)
The system has been put/kept under the specified environmental conditions.
(2)
Ambient temperature, relative humidity, and atmospheric pressure should meet the system requirements .
(3)
There should be no condensation .
(4)
There should be no deformation, damage, or dirt on the appearance .
• Clean the system when there is any dirt . Refer to "7 .2 Maintenance Check and Cleaning" and "8 .1 .6 Cleaning and
(5)
There should be no wobbling or loosening of screws in the casters, transducer holder, or peripheral devices, when the
LCD monitor and the Pole Cart2 are used .
(6)
The caster locks should function normally when the Pole Cart2 is used .
(7)
There should be no damage to the power supply cord, and/or loosened connectors .
(8)
There should be no dirt and abnormality (projection, cracking, chasm, sharp edge or peeling) on the appearance of the
transducer, cable, or connector .
• Clean and disinfect the transducer . Refer to "7 .2 .2 Cleaning Procedures" and "8 .1 .6 Cleaning and Disinfection" .
(9)
There should be nothing blocking the air vent of the system and no obstacles near the transducer cable, the wall mount-
ed AC socket outlet and mains plug .
(10)
The system should be placed in a stable position during operation .
(11)
There should be no obstacles that may interrupt operation, such as clips .
7 .1 .2 Check after Turning ON the Power
Check the items below after turning ON the power .
No .
Item
(1)
The transducer displayed on the system must match the connected transducer .
(2)
There must be no abnormal noise, abnormal smell, or heating .
(3)
No error message must be displayed .
(4)
There must be no intermittent display, or dark areas on the ultrasound image .
Refer to "8 .1 .4 Daily Inspection of Transducer" .
(5)
The date and time must be indicated correctly .
(6)
The acoustic lens surface of the transducer must not become abnormally hot . (Check by touching the lens surface .)
(7)
The buttons and dials on the mini console and the touch panel must function normally .
(8)
The operation sound must not be overly loud .
Refer to "System Screen" of the "Operation Manual: Applications" .
7 .1 Maintenance Checks Before and After Use
Summary of Contents for SONIMAGE HS2
Page 1: ...EN 01...
Page 2: ......
Page 6: ...6...
Page 7: ...7 Introduction...
Page 41: ...41 For all product inquiries contact your service representative Support...
Page 42: ...42...
Page 43: ...43 Chapter 1 Product Overview...
Page 68: ...68...
Page 69: ...69 Chapter 2 Before Examination...
Page 84: ...84...
Page 85: ...85 Chapter 3 Starting Examination...
Page 122: ...122...
Page 123: ...123 Chapter 4 Diagnosis Mode...
Page 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Page 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Page 232: ...232...
Page 233: ...233 Chapter 7 Maintenance Inspection...
Page 243: ...243 Chapter 8 Transducer and Biopsy...
Page 270: ...270...
Page 271: ...271 Chapter 9 MI TI...
Page 283: ...283 Chapter 10 Acoustic Output...
Page 286: ...286...
Page 287: ...287 Chapter 11 Others...
Page 305: ...305...
Page 306: ...306...
Page 307: ......