![Konica Minolta SONIMAGE HS2 Operation Manual Download Page 168](http://html1.mh-extra.com/html/konica-minolta/sonimage-hs2/sonimage-hs2_operation-manual_1994056168.webp)
168
The ECG function is a function that allows you to overlay the bright line of an ECG (electrocardiogram) on an ultra-
sound image, as well as synchronizing a B image with the bright line of the ECG for display .
To use this function, an optional "Reference Signal Unit" is required . Contact your service representative .
WARNING
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•
To prevent the electric shock, be sure to confirm the following cautions before use.
– There should be no abnormality on the ECG cable .
–
ECG Unit should be connected to the specified ECG cable for the Reference Signal Unit.
• Connect the Reference Signal Unit to the system prior to attach the ECG electrodes to a patient . Also, avoid bringing
the patient with the ECG electrodes attached into contact with other conductive parts including grounding . Doing so
may cause an electric shock to the patient when the connector of the ECG cable touches the live part .
• Do not connect the Reference Signal Unit to any connector other than USB connector . Doing so may cause a break-
down .
• Do not let ECG electrodes to touch patient’s heart directly . Doing so may cause a cardiac arrest .
– ECG is Type BF applied part, and is not intended for direct cardiac application .
• Be sure to use the ECG cable that is included with the Reference Signal Unit .
• The selection and use of ECG electrodes are the responsibility of a user .
• Remove any ECG electrodes from your patient before using equipment such as an electric scalpel, high-frequency
therapy equipment, electric stimulator or electric defibrillator. Also, make sure to keep any ultrasound transducer
away from the patient . The patient may get burned or receive an electric shock .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
CAUTION
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
• ECG display is information that is used strictly for your reference . Do not use these waveforms and the display for
diagnostic or monitoring purposes . Also, note that the heart rate of your patient may not be displayed correctly if he/
she uses a pacemaker or an implantable electric stimulator, or has an arrhythmia .
• Before using the Reference Signal Unit, check cables for damage, and the connector for looseness .
• Clean the ECG cable periodically .
– Be careful not to get the ECG clips wet . ECG waveforms will not be correctly displayed .
– Do not soak the connector section in water . Simply wipe it clean with a dry cloth . Entry of water into the connector can
cause a failure .
–
Wipe off any dirt on the ECG cable, using a cloth dampened with a solution of isopropyl alcohol diluted down to 30 % to
50 % or 70 % of ethyl alcohol .
•
Wipe off any dirt on the ECG Unit and USB cable using a soft cloth moistened with a mild detergent.
• The Reference Signal Unit is intended to connect to a medical device complying with IEC 60601-1 . Connect the Ref-
erence Signal Unit only to this system .
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
4 .7 ECG Mode
Summary of Contents for SONIMAGE HS2
Page 1: ...EN 01...
Page 2: ......
Page 6: ...6...
Page 7: ...7 Introduction...
Page 41: ...41 For all product inquiries contact your service representative Support...
Page 42: ...42...
Page 43: ...43 Chapter 1 Product Overview...
Page 68: ...68...
Page 69: ...69 Chapter 2 Before Examination...
Page 84: ...84...
Page 85: ...85 Chapter 3 Starting Examination...
Page 122: ...122...
Page 123: ...123 Chapter 4 Diagnosis Mode...
Page 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Page 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Page 232: ...232...
Page 233: ...233 Chapter 7 Maintenance Inspection...
Page 243: ...243 Chapter 8 Transducer and Biopsy...
Page 270: ...270...
Page 271: ...271 Chapter 9 MI TI...
Page 283: ...283 Chapter 10 Acoustic Output...
Page 286: ...286...
Page 287: ...287 Chapter 11 Others...
Page 305: ...305...
Page 306: ...306...
Page 307: ......