418
12 .1
Specifications
12 .1 .1
Software specifications
Item
Description
Product name
DIRECT DIGITIZER CS-7
Product classification
MDD: Class
II
a, Rule10
FDA: 510 (k) Class
II
Remarks (for MDD)
CE marking, indicating conformity with the Medical Devices Directive 93/42/EEC
DICOM support
DICOM Storage SCU
DICOM Basic Grayscale Print Management SCU
DICOM Modality Worklist Management SCU
DICOM Modality Performed Procedure Step SCU
Image processing
Auto-gradation process
Frequency processing (F process)
Equalization processing (E process)
Hybrid processing (HF process - HE process)
Hybrid smoothing process (HS process)
Grid removal process/Scattered Radiation Correction (Intelligent-Grid)
Automatic exposure field recognition process
Readable devices
1) DR Detector
2) REGIUS MODEL 170, REGIUS MODEL 190, REGIUS MODEL 210, REGIUS MODEL 110,
REGIUS MODEL 110S, REGIUS MODEL 110HQ
3) REGIUS SIGMA, REGIUS SIGMA2
Sampling pitch
200μm (Only for AeroDR 3 1417HD)
175μm
100μm (Only for AeroDR 3 1417HD)
87.5μm (Only for REGIUS series)
Image output
Host: max 4ch
Printer: max 2ch
Software components
1) Application : CS-7 application software
2) OS :
Windows 7 Professional 32bit/Windows 10 Enterprise 2015 LTSB (64bit)
------ SOUP
3) Middleware : .NET framework 4.6.1
------------------------------------------------------------------------ SOUP
Microsoft SQL Server Compact Edition -------------------------------------------------- SOUP
Postgre SQL ------------------------------------------------------------------------------------- SOUP
4) Driver :
CD-R/DVD-R writing library
----------------------------------------------------------------- SOUP
* OS is installed in PC . It is not included in CS-7 product .
Software main functions
1) Order registration/Order cooperation
The order registration method of this device can be selected according to the connection environment
of the facility .
2) Exam. Tag
Select the Exam. Tag for exposure to select the optimal reading condition, image-processing condi
-
tion, and output condition for each body part .
3) Image reading
4) Linkage with X-ray equipment
Linkage of exposure between X-ray equipment, Digital Radiography, and this device.
5) Image processing
6) Image sending to printer/PACS/image storage media
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•
This device does not have a function for controlling the X-ray device. Exposure is operated via the X-ray device.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...