7 .18 Checkup mode
4
Turn [Daily Reset] and [Auto Checkup]
[ON] and [OFF], as necessary.
5
Select the preset used .
6
Press [OK] .
• The display returns to the system monitor
screen .
7
Press [Close] .
• The display returns to the initial screen .
• When serial numbers are automatically issued,
the list screen or the patient registration screen
is displayed. Press [Perform] to start the expo
-
sure in checkup mode .
7 .18 .4 Exposure in checkup
mode
The following describes the procedure for performing
exposures in the checkup mode.
Before starting the exposure, refer to
checkup mode" to make necessary preparations for
the checkup mode .
The following procedures vary depending on whether
the auto examination start function is to be used, and
also whether the Exam. Tags are already set for the
preset used .
z
When the Auto Checkup is turned ON and
the Exam . Tags are set
The following describes the procedure when [Auto
Checkup] is turned [ON] and the Exam. Tags are set
for the preset used .
1
Display the Patient registration screen or
List screen.
• Check that the automatic issuing worked cor-
rectly .
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...