236
7.4 Exposure with Unitless
7 .4 .3 Exposure
Prepare for exposure using unitless according to the
following procedure . The procedure varies depending
on the setting of the [AeroSync Button] on the system
monitor screen .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
Always check the patient information and exposure
conditions before exposure.
• Do not disconnect the Ethernet cable while the data is
being transmitted .
• The status of communication with the host is displayed
on the [System] button. For detail, refer to
• Available X-ray detection level varies depending on
the DR Detector model .
• Use the PDA at a distance within 10 m of this device
and the DR Detector .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
•
The procedure is explained with the use of the Aero
-
DR 3 1417HD as an example.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is OFF
1
After arriving to the imaging location, con-
firm that the icon for the DR Detector used
for imaging appears in the application bar .
2
Select the examination .
3
Press [Perform] .
•
The exposure screen is displayed.
• A message "Ready" is displayed in the mes-
sage/button area, then the screen is locked .
• To change the X-ray detection level, press
[Pause] to enable modification. After changing
the detection level, press [Restart], then the
message "Ready" is displayed .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• A malfunction may be caused by a shock to the DR
Detector. In such cases, set [AeroSync Button] to ON
on the system monitor screen .
• To prevent incorrect operations, always check that
"Ready" is displayed .
• Conditions for changing the detection level vary de-
pending on the X-ray device being used . If a survey
has been carried out when the system was installed,
follow the survey results .
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
•
When importing examination information from the
higher system and when exposure body part information
is not imported, select the exposure body part. When
pressing [Perform] on the body part selection screen, the
exposure screen is displayed.
• When you move the focus in the image display area on
the exposure resumption waiting screen to capture an im
-
age of another body part, not the ready-to-expose screen
but the exposure resumption waiting screen is displayed.
• Refer to
for information on
the X-ray detection levels that can be set for exposures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...