Setting Measurement Data as Target
Set the measurement data as the target data.
Display the sample to be set as the target on the <Sample> screen in advance.
Setting Procedure
Start the procedure from the <Sample menu> - <Data management> screen.
Memo
•
Target colors are stored with setting numbers from 0001 to 0200 being assigned. Even when the data at some point
is deleted, these setting numbers do not change.
•
To set more accurate target color data, use averaging to measure the target specimen. For details, refer to the
“Average Measurement” section on pages 37 and 69 to 71.
Notes
• Be sure to perform white calibration before setting target colors.
• To ensure measurement accuracy, make sure to keep ambient conditions (temperature, etc.) constant.
• Unlike measured data, the numbers assigned to target color data do not change automatically. When
measuring colors continuously to set target colors, the cursor must be moved manually to set each data.
1
Use [
] or [
] to move the cursor to “Set Sample as
Target”, and then press the [Confirmation] key.
The <Set Sample as Target> screen
is displayed.
2
The target color data number being used will be
highlighted. Use [
] or [
] to move the cursor to
the color difference reference color number, and
then press the [Confirmation] key. When the target
color data is set, the screen returns to the <Target>
screen.
•
If
[ESC]
is pressed without pressing the
[Confirmation]
key
, the screen will return to the
<Data management>
screen
.
Notes
If a number to which target color data has already been
set is selected, a message is displayed to confirm
overwriting. Use [
] or [
] to select “OK”. Press the
[Confirmation] key. Moving the cursor to “Cancel” and
pressing the [Confirmation] key will cancel the deletion,
and the screen will return to the <Target> screen.
Data management
Sample
Sample menu [0013]
0002 TARGET-2
Delete data
Change list position
Change Target reference
0002 TARGET-2
Set Sample as Target
Data management
Sample
Sample menu [0013]
0002 TARGET-2
Delete data
Change list position
Change Target reference
0002 TARGET-2
Set Sample as Target
Set Sample as Target
/ 2 0 0
0 0
2
2016/08/23 16:27:53
0002: TARGET-2
Data management
Sample
Sample menu [0013]
0002 TARGET-2
Delete data
Change list position
Change Target reference
0002 TARGET-2
Set Sample as Target
Warning
OK
Cancel
Are you sure you want to
overwrite all targets?
WARNING
Summary of Contents for CM-M6
Page 1: ...Instruction Manual Please read before using the instrument...
Page 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Page 102: ...99...
Page 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Page 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Page 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Page 126: ...123...
Page 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...