E-80
Measuring the Average
Auto Averaging
Measurement is repeated the specified number of times in the same position on the specimen and then
the average of the measured reflectance data is calculated. This will improve the accuracy of the meas-
ured data.
Set the number of measurements to be performed and the standard deviation, and then start measurement.
N
ote
Before starting auto averaging, make sure that the number of measurements to be performed is set. For details,
refer to “Setting a Measurement Condition” (from page E-34).
[Setting Procedure]
This operation must be started from a screen in which “BREAK” is displayed.
1
Make sure that “BREAK” is highlighted, then press
A
.
Turn
B
to select “BREAK”,
if it
is not highlighted.
• Example when “COND1” is selected
2
Place the speimen measuring port to the specimen, and then press
C
(MEAS.
button).
• Take care to keep the instrument still during measurement.
The number of lamp flashing will be
highlighted during auto averaging.
When the specified number of meas-
urements has been completed, the
averaging result will be displayed on
the LCD.
M
emo
If
does not appear due to the
auto power save function, this may
be due the fact that it can take time to
start measurement after
C
is
pressed.
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...