Change Target Reference
The target data acting as a reference for the sample can be changed.
Display the sample for which the target reference is to be changed on the <Sample> screen in advance.
Setting Procedure
Start the procedure from the <Sample menu> - <Data management> screen.
1
Use [
▲
] or [
▼
] to move the cursor to “Change Target
reference”, and then press the [Confirmation] key.
The <Change Target reference> screen
is displayed.
Notes
• If a target with different measurement conditions is
specified, an error message will be displayed. Change
to a target with the same measurement conditions.
2
The target data number being used will be
displayed. Use [
▲
] or [
▼
] to move the cursor to the
color difference target number, and then press the
[Confirmation] key. When the target data is set, the
screen returns to the <Sample> screen.
• If
[ESC]
is pressed, the target changes will not be
applied and the screen will return to the
<Data
management> screen
.
Notes
• If a number to which no color difference target data
has been set, the target will be determined as “None”
and the color difference and pass/fail judgment will
not be displayed.
Summary of Contents for CM-23d
Page 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Page 30: ...27...
Page 82: ...79...
Page 111: ...108 Setting 3 MEMO...
Page 112: ...109...
Page 132: ...129...
Page 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Page 138: ...135...
Page 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Page 142: ...139 Dimensions Unit mm...
Page 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...