52
6�1
Specifications
IMPORTANT
•
The following specifications described below are nominal values which may be different from actual values and may vary de-
pending on environment and frequency of use. (These are not to provide any guarantees.)
•
All specification regarding battery is for a fully-charged battery.
6�1�1
DR Detector
Item
Description
Product name
P-41
Detection method
Indirect conversion method
Scintillator
CsI (Cesium Iodide)
Power supply
AC Adapter : DC 24 V 0.75 A
Battery
: DC 11.5 V 1.5 A
External dimensions
384(W)×460(D)×15(H) mm
460 mm
15 mm
384 mm
Weight
3.6 kg (Including the Battery)
Pixel size
150 um
Image area size
345.6×420 mm (2,304×2,800 pixels)
AD conversion
16 bit
Maximum patient weight
Point load:100 kg@Φ40 mm
Face load:150 kg@effective image area overall
* Deadweight, even when loaded on the DR Detector, has no effect on images and the DR
Detector. The measurement method is based on KM standards.
Communication
Dedicated wired Ethernet connection / wireless LAN (IEEE802.11a/b/g/n/ac compliant)
* Dedicated wired Ethernet connection is reserved for service maintenance
Drive power
Dedicated power cable drive / battery drive
WLAN encryption
Wireless encryption method : AES
Authentication method
: WPA2-PSK
Wireless QoS
Should be provided with standard-compliant
Patient doses
Patient doses are equivalent to 500-1000 speed film/screen systems.
Summary of Contents for AeroDR SKR 4000
Page 1: ...EN 02...
Page 2: ......
Page 8: ...8...
Page 9: ...9 Chapter 1 Safety Precautions Warnings...
Page 22: ...22...
Page 23: ...23 Chapter 2 Product Overview...
Page 29: ...29 Chapter 3 General Operations...
Page 42: ...42...
Page 43: ...43 Chapter 4 Troubleshooting...
Page 46: ...46...
Page 47: ...47 Chapter 5 Maintenance...
Page 50: ...50...
Page 51: ...51 Chapter 6 Specifications...
Page 56: ...56...
Page 57: ...57...
Page 58: ...58...
Page 59: ......