Chapter 5 Disassembly and Assembly
5-118
5.10.3 Replacing the CF Card
This section describes the instructions on how to replace the CF card.
Before Replacing the CF Card
Own various configuration data on the individual REGIUS MODEL 110 equipment is stored in each CF
card. Therefore, data within the card must be backed up before replaced, then it must be restored in the
new card after replaced. The service tools can be used to perform these works as above on the REGIUS
console that is connected to the REGIUS MODEL 110 through networking.
Work Procedures
This section describes the procedures for the works required in order to replace the CF card.
For operating the service tools, see the Installation/Service Manual for REGIUS console.
Work outline
Listed below are the number of people and the rough standard of work hours required for performing the
works described in this section.
Requirements
Shown below are the tools requiring to be prepared in addition to the standard tools before performing the
works in this section.
*: Cross cable can also be used.
1
If the equipment can communicate with the REGIUS console, back up data in the CF card and
store it in the console in the "Change Circuit Board" screen of a service tool.
2
Back up the own configuration data (system or JM information) on the REGIUS on the console
on which backup of the CF card data has been performed.
Subsequently, use this REGIUS console to restore the CF card data or configure the equipment (REGIUS
MODEL 110).
3
Reset the settings in the REGIUS console to default (1:1 connect, post-register)
Network address setting screen
•
IP address: 192.168.20.90
•
Subnet mask: 255.255.255.0
JOBINFO screen
•
Select "JobManager internal, post-register"
Important
Replace the CF card when it is possibly malfunctioning. In most cases, a malfunction in the CF
card will disable the REGIUS MODEL 110 to establish network connection, resulting in
impossibility to perform data backup at this time. The CF card backup is one of the work items
required when installing the equipment, however, backup must be performed whenever any
setting is changed in the REGIUS MODEL 110 for the purpose of servicing or other purpose.
Caution
•
Be sure to turn off the power supply circuit breaker and unplug the power cable from the
equipment whenever you disassemble/assemble the equipment.
•
See
"5.2.7 Binding the Grounding Strap (Page 5-18)"
to bind a grounding strap in order to
protect the circuit board from damage due to static discharge.
Personnel
Number
Work Hours
1
No.
Tool
No.
Tool
1*
Ethernet cable
(Category 5e or higher)
2*
Network hub
Summary of Contents for 0902
Page 2: ......
Page 6: ......
Page 12: ......
Page 30: ...2 10...
Page 54: ......
Page 84: ...Chapter 3 Troubleshooting 3 30 3 2 19 Confirm the H sync Signal To be stated...
Page 85: ...Chapter 3 Troubleshooting 3 31 3 2 20 Confirm the LMC Operation To be stated...
Page 120: ......
Page 296: ...8 10...
Page 300: ......
Page 301: ......