476671/B
323
Result
Requirements
Results
The information presented by the
B-Scan
page verifies that the MF90 system is
fully operational.
All elements are operational in both transmit and receive. The maximum
number of bad elements is: 3
Weak elements:
Unserviceable elements:
Date and signature:
Measuring the noise in passive mode
Low noise is a key factor for high quality and reliable measurements. The performance of
the MF90 system will always be limited by different noise sources. A high noise level will
mask the echoes and make them difficult to read and interpret. The noise is measured while
the MF90 system operates in
Passive
mode with the transmit pulses disabled.
Prerequisites
The MF90 system is turned on and operates normally.
• The MF90 system is in
Normal
mode, but
TX Power
is set to
Off
to prevent transmissions.
• The transducer is retracted to its upper position.
• Neither tools nor instruments are required.
In order to do this test, the ship must be "silent".
• The water must be as deep as possible. Recommended minimum depth is 100 metres.
• There must be no other vessels in the vicinity.
• The vessel must lie still in the water.
• As much machinery as possible must be turned off. It is particularly important to turn
off electrical motors, as well as cooling systems and hydraulic pumps that may cause
electric noise.
• To prevent interference, all other hydroacoustic instruments must be turned off.
Note
With the vessel in port, the environmental conditions are not satisfactory. In the shallow
waters of the port, noise from other vessels, dockyard workers or machinery will cause
unreliable test results. If you do this tests in a busy harbour, or with noise sources present,
the sensitive receivers will detect all the noise in the nearby waters.
Summary of Contents for Simrad MF90
Page 1: ...kongsberg com simrad Simrad MF90 INSTALLATION MANUAL...
Page 2: ......
Page 21: ...476671 B 19 Related topics Simrad MF90 page 13 Simrad MF90...
Page 369: ...476671 B 367 Drawing file...
Page 372: ...370 476671 B Simrad MF90 Installation Manual...
Page 375: ...476671 B 373 Drawing file...
Page 378: ...376 476671 B Simrad MF90 Installation Manual...
Page 384: ...382 476671 B Simrad MF90 Installation Manual...
Page 387: ...476671 B 385 Drawing file...
Page 390: ...388 476671 B Simrad MF90 Installation Manual...
Page 393: ...476671 B 391 Drawing file...
Page 398: ...396 476671 B Simrad MF90 Installation Manual...
Page 421: ......
Page 422: ...2022 Kongsberg Maritime ISBN 978 82 8066 240 8...
Page 423: ......
Page 425: ......
Page 426: ...Installation Manual Simrad MF90...