KV4000 Kinematic Viscosity Bath
Operation and Technical Manual
KV4000-Manual
-
19 -
7.1 Unit does not power-up
Establish that the socket outlet is providing
proper and adequate voltage.
Check if Overpower Protection circuitry
located directly behind the temperature
controller inside the front tray has been
activated.
Check if line switch is in the
ON
position.
If problem persists, please call the Koehler
technical service department for assistance.
7.2 Unit is on and keeps resetting into
start up routine
For 220V units, make sure that the socket
outlet is greater than 215V.
Check if there is a steady and reliable power
source.
Make sure the connector plug on the rear
panel is firmly plugged in.
7.3 Unit is on but bath does not heat up
Make sure that the actual temperature
reading is not higher than the set-point
temperature.
Determine if the temperature controller is
properly calibrated by comparison to an
ASTM standard thermometer.
Determine if the Overtemperature Protection
(OTP) circuitry as been activated.
7.4 Bath heats up but temperature does
not stabilize
Make sure there are no drafts from open
doors, windows, or environmental control
vents in the vicinity of the bath.
If the set temperature is close to ambient,
then you may need to circulate cold water
through the coils in order to achieve proper
temperature stability
.
8
Maintenance
WARNING:
Disconnect power to the unit
before servicing to avoid exposure to high
voltages and/or temperatures which may result
in personal injury or death. If you have any
questions about maintaining your equipment,
then please do not hesitate to contact the
Koehler technical service department.
8.1 Routine Maintenance
The KV4000 Kinematic Viscosity System
requires little routine maintenance to provide
many years of continuous service. However,
over the course of time, some instrument parts
may need to be replaced. When ordering
replacement part(s), please provide the model
number, serial number, and product shipment
date of your equipment so that we can ensure
you will receive the proper replacement part(s).
8.2 Bath Cleaning
To clean
the instrument’s exterior, which
includes all painted surfaces and glass,
either a solution of soap and water or
laboratory grade detergent may be used.
Apply cleaner to clean wipe or cloth, not to
the instrument directly. Wipe surface clean.
Do Not
clean bath exterior with organic
chemicals such as Acetone, Toluene,
Hexane, etc.
For more difficult cleaning of finished
surfaces, a dilute solution or isopropanol in
water may be used.
It is not recommended that more aggressive
solvents be used on painted surfaces since
paint color will tarnish or be stripped from
the instrument.
Stainless Steel surfaces, such as on the top
plate, may be cleaned using a more
aggressive solvent such as a stainless steel
cleaner.
Glass surfaces may be cleaned using a
more aggressive solvent such as acetone, if
necessary.
Summary of Contents for KV4000
Page 2: ......
Page 4: ......
Page 5: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 1...
Page 6: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 2...
Page 29: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 25 Notes...
Page 30: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 26 Notes...