K430X2 Dual Twin Foaming Characteristics Test Bath 

Operation and Instruction Manual 

 

K430X2-Manual

    

 

-

14-

 

6  Safety Features

 

 

The  Koehler  K430X2  Dual  Twin  Foaming  Test 

Bath  is  equipped  with  several  safety  and 
protection  features,  which  are  described  in  the 
following sections. 

 

WARNING:

  The  bath  is 

NOT

  Explosion 

Proof.  

 

6.1  Over-Temperature Protection

 

 
The  Koehler  K430X2  Dual  Twin  Foaming  Test 

Bath is equipped with Over-temperature Protection 
(OTP)  circuitry,  which  prevents  the  unit  from 
exceeding  unsafe  operating  temperatures.  If  the 
unit cannot maintain the set point temperature and 
begins  to  decline,  the  OTP  circuitry  may  have 
been activated. Please follow these steps. 

 

1.  Turn off the unit  power switch and disconnect 

the line cord. 
 

2.  Determine  the  source  of  the  problem  and 

correct the situation. 
 

3.  Restart  the  unit.  Monitor  the  operations  to 

ensure  that  the  unit  is  operating  properly.  If 
you  are  still  experiencing  trouble,  please 
contact 

Koehler 

technical 

service 

for 

assistance. 

 

6.2  Over-Power Protection

 

 
The  Koehler  K430X2  Dual  Twin  Foaming  Test 

Bath  is  equipped  with  Over-power  Protection 
circuitry,  which  prevents  the  unit  from  unsafe 
electrical  conditions.  If  power  to  the  unit  is  lost, 
then  turn  off  the  main  power  and  turn  it  back  on 
again. The main power switch also functions  as a 
circuit breaker. 

 

 
 
 
 
 
 
 
 
 
 

7  Maintenance

 

 

WARNING.

 

Disconnect  power  to  the  unit 

before  servicing  to  avoid  exposure  to  high 
voltages  and/or  temperatures  which  may  result  in 
personal injury or death. If you have any questions 
about maintaining your equipment, then please do 
not  hesitate  to  contact  the  Koehler  technical 
service department. 
 

7.1  Routine Maintenance

 

 

The  K430X2  Dual  Twin  Foaming  Test  Bath 

requires little routine maintenance to provide many 
years  of  continuous  service.  However,  over  the 
course  of  time,  some  instrument  parts  may  need 
to  be  replaced.  When  ordering  replacement 
part(s),  please  provide  the  model  number,  serial 
number,  and  product  shipment  date  of  your 
equipment so that  we can  ensure  you  will receive 
the proper replacement part(s). 
 

7.2  Bath Cleaning 

 

 

To  clean  the  instrument’s  exterior,  which 
includes all painted surfaces and glass, either 
a  solution  of  soap  and  water  or  laboratory 
grade detergent may be used.  

 

  Apply cleaner to clean wipe or cloth, not to the 

instrument directly. Wipe surface clean. 

 

 

Do  Not

  clean  bath  exterior  with  organic 

chemicals such as Acetone, Toluene, Hexane, 
etc.  

 

  For more difficult cleaning of finished surfaces, 

a  dilute  solution  or  isopropanol  in  water  may 
be used. 

 

  It  is  not  recommended  that  more  aggressive 

solvents  be  used  on  painted  surfaces  since 
paint color  will  tarnish  or  be  stripped  from  the 
instrument.  

 

  Glass  surfaces may  be  cleaned  using  a  more 

aggressive  solvent  such  as  acetone,  if 
necessary.  

 

SHOCK  AND  BURN  HAZARD:

  Only  clean 

inside  the  bath  when  equipment  is  de-energized 
and  unplugged  from  the  mains  power  supply. 
Allow adequate time for heating coils to completely 
cool before cleaning.  

 

Summary of Contents for K430X2

Page 1: ...ohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K430X2 Dual Twin Foaming Characteristics Test Apparatus ...

Page 2: ......

Page 3: ...with the applicable standards set forth in this certification Specifications ASTM D892 IP 146 DIN 51566 FTM 791 3211 FTM 791 3213 NF T 60 129 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrument Company ...

Page 4: ......

Page 5: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 1 ...

Page 6: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 2 ...

Page 7: ...asic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is b...

Page 8: ...eparately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbo...

Page 9: ... Flowmeter Setup and Operation 11 4 3 Flowmeter Calibration Data 11 4 4 Temperature Controller Operation 11 5 Operation 13 5 1 Bath 13 5 2 Power 13 5 3 Preparation of Glassware and Tubing 13 6 Safety Features 14 6 1 Over Temperature Protection 14 6 2 Over Power Protection 14 7 Maintenance 14 7 1 Routine Maintenance 14 7 2 Bath Cleaning 14 7 3 Replacement Parts 15 8 Wiring Diagrams 16 8 1 Unit Wiri...

Page 10: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 6 12 Warranty 17 13 Returned Goods Policy 18 Notes 19 ...

Page 11: ...specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests ...

Page 12: ... and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certifications whether specified or implied and may result in personal injury and or property loss Replacement parts must be O E M exact replacement equipment Over Temperature Protection This ...

Page 13: ...rom box to cart and from cart to bench Individuals must lift in accordance to proper technique See Figure below 7 Lift instrument from cart and place on bench 8 Ensure that all parts listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your cl...

Page 14: ...1 Main Instrument Controls Figure 1 Main Instrument Descriptions 1 Power Switch This switch controls the power to motor and heater temperature controller of its corresponding side When the power switch is in the ON position the digital temperature controller and the stirrer are powered on 2 Temperature Controller The temperature controller regulates the bath temperature its corresponding bath Refe...

Page 15: ...the duration specified in the test procedure and disconnect the tygon tubing from the flowmeter outlet when the aeration period as elapsed 4 3 Flowmeter Calibration Data K43002 0 12 Air Flow Rate 94 mL min Float Material Sapphire Float Density 3 98 g ml Metered Fluid Air Minimum Flow 16 4 ml min Maximum Flow 159 0 ml min Tolerance 2 F S Scale Reading mm Flowrate ml min 65 159 0 60 139 0 55 121 2 5...

Page 16: ...is a value observed in the upper red LED display add it to the calculated difference obtained in the previous step This is the offset value c Press the Program Key The lower green display will flash Use the up or down keys to adjust to the new calibration offset value on the upper red display calculated in the previous step When the value has been entered the controller will automatically store th...

Page 17: ...s that will decompose or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special waste...

Page 18: ...e servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 7 1 Routine Maintenance The K430X2 Dual Twin Foaming Test Bath requires little routine maintenance to provide many years of continuous service Howeve...

Page 19: ...0Hz 7A 278 001 002 Fuse 1A Slo Blo 5 X 20 mm 332 001 003 12 x 18 Borosilicate Bath Jar K23425 03000 18 Safety Shield K23700 03014A Motor 230V 50 60HZ 1 15HP K430 0 13 Air Outlet Elbow Tube K430 0 8 Rubber Stopper K430 3A Air Inlet Delivery Tube without diffuser K43002 0 12 Low Flowmeter K43002 0 13 Test Tube Rack K43002 0 15 Locking Post Assembly K43012 1000ml Cylinder w Ring K43002 0 9 Hot Bath O...

Page 20: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 16 8 Wiring Diagrams 8 1 Unit Wiring ...

Page 21: ... Dual Twin Foaming Characteristics Test Bath should not require service Any service problem can be quickly resolved by contacting Koehler s technical service department either by letter phone fax or email In order to assure the fastest possible service please provide us with the following information Model Number Serial Number Date of Shipment 11 Storage This laboratory test instrument is equipped...

Page 22: ...er the item s to be returned and the reason for the return You will be issued a Returned Authorization RA number which must be prominently displayed on the shipping container when you return the material to our plant Shipping containers without an RA number prominently displayed with will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charg...

Page 23: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 19 Notes ...

Page 24: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 20 Notes ...

Reviews: